Failure to detect life-threatening arrhythmias in ICDs using single-chamber detection criteria
Roland X Stroobandt1, Mattias F Duytschaever1,2, Teresa Strisciuglio2,3
1Heart Center, Ghent University Hospital, Ghent, Belgium.
Background:
There are anecdotal reports of sudden death despite a functional implantable cardioverter defibrillator (ICD). We sought to describe scenarios leading to fatal or near-fatal outcome due to inappropriately inhibited ICD therapy in devices programmed with single-chamber detection criteria.
Methods:
Programmed settings, episode lists, and intracardiac electrograms from 24 patients with a life-threatening event (n = 12) or fatal outcome (n = 12) related to failed ventricular arrhythmia detection were used to clarify the underlying scenario.
Results:
Fifty episodes of failed ventricular arrhythmia detection were identified and categorized into six scenarios: (1) spontaneous ventricular tachycardia (VT) or ventricular fibrillation (VF) with a rate below the detection limits, (2) misclassification of polymorphic VT (PVT) or VF as supraventricular tachycardia (SVT), (3) misclassification of VT/VF as cluster of nonsustained VT episodes, (4) misclassification of monomorphic VT (MVT) as SVT, (5) inappropriate shock abortion, and (6) false termination detection. These scenarios occurred respectively 6, 9, 3, 9, 8, and 15 times. In 9/9 (100%) patients with PVT/VF classified as SVT, rate stability was active for rates ranging from 222 to 250 beats/min. MVT detected as SVT was due to the sudden onset criterion in 7/9 (78%) patients and twice a consequence of the rate stability criterion active for rates ranging from 200 to 250 beats/min.
Conclusion:
We describe six scenarios leading to failure of ventricular arrhythmia detection in a single-chamber detection setting withholding life-saving therapy. These scenarios are more likely to occur with high-rate programming and long detection times, especially if combined with rate stability and sudden onset.
More Related Videos
11:27A Modified Sonographic Algorithm for Image Acquisition in Life-Threatening Emergencies in the Critically Ill Newborn
Published on: April 7, 2023
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Criteria for Causality: Bradford Hill Criteria - II
Criteria for Causality: Bradford Hill Criteria - I
Half-life of a Reaction
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Characteristics of Life