Initial exploration of coronary stent image subtraction using dual-layer spectral CT
Le Qin1, ShengJia Gu1, ChiHua Chen1
1Department of Radiology, Ruijin Hospital Affiliated to Shanghai Jiaotong University School of Medicine, No. 197 Ruijin 2nd Rd, Shanghai, 200025, China.
Objectives:
This study aimed to investigate the feasibility of coronary stent image subtraction using spectral tools derived from dual-layer spectral computed tomography (CT).
Methods:
Forty-three patients (65 stents) who underwent coronary CT angiography using dual-layer spectral CT were included. Conventional, 50-keV (kilo electron-volt), 100-keV, and virtual non-contrast (VNC) images were reconstructed from the same cardiac phase. Stents were subtracted on VNC images from conventional (convsub), 100-keV (100-keVsub), and 50-keV (50-keVsub) images. The in-stent lumen diameters were measured on subtraction, conventional, and 100-keV images. Subjective evaluation of reader confidence and subtractive quality was evaluated. Friedman tests were performed to compare in-stent lumen diameters and subjective evaluation among different images. Correlation between stent diameter and subjective evaluation was expressed as Spearman's rank correlation coefficient (rs). The diagnostic accuracy was assessed according to invasive coronary angiography (ICA) performed in 11 patients (20 stents).
Results:
In-stent lumen diameters were significantly larger on subtraction images than those on conventional and 100-keV images (p < 0.05). Higher reader confidence was found on 100-keV, convsub, 100-keVsub, and 50-keVsub images compared with conventional images (p < 0.05). Subtractive quality of 100-keVsub images was better than that of convsub images (p = 0.037). A moderate-to-strong correlation between stent diameter and subjective evaluation was found (rs = 0.527~0.790, p < 0.05). Higher specificity, positive predictive value, and negative predictive value of subtraction images were shown by ICA results.
Conclusions:
Subtraction images derived from dual-layer spectral CT enhanced in-stent lumen visibility and could potentially improve diagnostic performance for evaluating coronary stents.
Key Points:
• Dual-layer spectral CT enabled good subtractive quality of coronary stents without misregistration artifacts. • Subtraction images could improve in-stent lumen visibility. • Reader confidence and diagnostic performance were enhanced with subtraction images.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Transcription Initiation
The promoters and enhancers and their accessory proteins allow tight regulation of...