Related Experiment Video
Updated: Jan 30, 2026

Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
A sensitive and robust UPLC-MS/MS method for quantitation of estrogens and progestogens in human serum
Junmei Zhang1, Chenxiao Tang1, Patrick J Oberly1
1Department of Pharmaceutical Sciences, School of Pharmacy, University of Pittsburgh, Pittsburgh, PA.
Objective:
With the widespread use of sex-steroid hormones in contraceptives and hormone replacement therapy, there is an increasing need for reliable analytical methods. We report the development of a sensitive and robust UPLC-MS/MS method for quantitation of both endogenous and synthetic sex-steroid hormones in human serum.
Study Design:
We developed and validated a UPLC-MS/MS method to quantify progestogens (etonogestrel, levonorgestrel, medroxyprogesterone acetate, norethindrone, progesterone) and estrogens (estradiol and ethinyl estradiol) with good accuracy, high sensitivity, and excellent robustness. We then applied the method to the analysis of sex-steroid hormones in serum from 451 clinical research participants.
Results:
Each UPLC-MS/MS analysis was 6.5 min. The lower limits of quantitation (LLOQs) were 25 pg/ml for the progestogens, and 2.5 and 5.0 pg/ml for estradiol and ethinyl estradiol, respectively. When estradiol was analyzed without assessment of progestogens, the LLOQ was reduced to 1 pg/ml. The calibration curves were linear from 25-50,000, 2.5-2000 (1-2000 for estrogens-only analysis) and 5-2000 pg/ml, respectively. Both the accuracy and precision were below±15% not only for routine validation (intraday and interday), but for long-term (>2 years) assay robustness with external controls, thereby, demonstrating the utility of this method for multi-year clinical trial assessments of progestogens and estrogens. We applied the method to quantify sex-steroid levels in 1804 clinical samples.
Conclusions:
We successfully developed a UPLC-MS/MS method, and overcame the matrix suppression to allow sensitive quantitation of both synthetic and endogenous sex-steroid hormones in human serum.
Implications:
We developed a sensitive and robust UPLC-MS/MS method to accurately measure the levels of sex-steroid hormones in serum. The method overcame matrix interference barriers and achieved excellent long-term stability and reproducibility (≥96.9% accuracy; ≤13.0% relative variability measured with external controls over 2 years), demonstrating its utility in clinical sample analysis.
Related Concept Videos
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...
Serum Laboratory Studies, Stool Test, Breath Test
Serum Studies: Renal Function Tests
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
The Scientific Method