Piezoelectric Device Versus Conventional Osteotomy Instruments in the Comparison of Three Different Bone Harvesting
Purpose:
This study was conducted to compare the bone cutting performance of the piezoelectric device with traditional drills in terms of cut execution time, environmental noise produced, surface morphology characteristics, and residual cell vitality of the bone samples.
Materials And Methods:
Two fresh pig jaws were obtained from a local slaughterhouse, and nine bone samples with a standard size (1 cm3) were harvested from each jaw: three using piezosurgery, three with a multiblade bur mounted on a surgical handpiece, and three with a diamond bur mounted on a high-speed handpiece, for a total of 18 samples. Two samples for each harvesting method were examined by scanning electron microscope (SEM), observing four surfaces per sample. For each surface observed, a count of the intertrabecular spaces was performed, and each space was evaluated as completely, partially, or unfilled by debris. Four samples per sampling method were examined by cell culture to evaluate residual cell vitality after cutting. The execution time of each osteotomy was measured with a stopwatch. The environmental noise was measured, at two different distances, with a phonometer.
Results:
At SEM analysis, piezosurgery osteotomies showed 66% of totally free intertrabecular spaces on the cutting surface, statistically significantly higher than those on the multiblade bur (33%) and diamond bur (12%) cutting surfaces (P < .0005). Bone samples harvested with piezosurgery also demonstrated faster cell proliferation. Finally, piezosurgery generated less environmental noise, though it required longer cutting time.
Conclusion:
Piezoelectric technology is a valid alternative to rotating burs for osteotomy, demonstrating higher bone cell viability and a precise and silent, though slower, cut.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
10:39Fabrication and Characterization of Thickness Mode Piezoelectric Devices for Atomization and Acoustofluidics
Published on: August 5, 2020
Related Concept Videos
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Uncertainty in Measurement: Reading Instruments