10-year results of ReCap hip resurfacing arthroplasty: a non-designer case series
Manish Kiran1, Sunil Santhapuri1, Sara Moeen1
1Bedford Hospital, Bedford, UK.
Introduction:
We present the 10-year results of the ReCap Magnum HRA system in a consecutive single surgeon's series from a non-designer centre.
Patients And Methods:
Hip resurfacing arthroplasty (HRA) was performed in active males under 65 years and in pre-menopausal females with adequate bone density proven by a DEXA scan. Clinical follow-up; annual Oxford hip scoring, UCLA scoring and whole blood metal ion level measurement. Kaplan-Meier method was used to determine survivorship with revision as the end point. Radiographs were analysed for; acetabular inclination, notching, neck thinning and change in implant position. 72 HRAs were performed in 66 patients with a mean age of 45.7 years. The mean follow-up was 10.63 ± 0.54 years (range 10-11.75 years).
Results:
The survivorship of the cohort at minimum 10 years was 97.22% (95% confidence interval 94.14-99.01). There was a significant improvement in the Oxford Hip Score from 15.77 ± 4.33 to 45.67 ± 4.43 (p < 0.001) and the UCLA score from 3.14 ± 0.74 to 7.07 ± 1.16 (p < 0.001). Mean whole blood cobalt and chromium levels were 28.83 ± 8.42 nmol/l and 39.93 ± 9.64 nmol/l respectively. There were 2 revisions, due to avascular necrosis and implant failure. In our series the implant has performed well with <1% failure per year. We did not note any difference between the genders, since all the females in our series had been screened for osteopenia using a DEXA scan. All our patients returned to an active lifestyle which was maintained at final follow-up.
Conclusion:
Strict selection criteria and a well-designed implant result in good long-term functional and radiological results with hip resurfacing.
Related Concept Videos
Resistors In Series
In a series circuit, the...
Group Design
Series Resonance
Series and Parallel Capacitors
First, consider capacitors connected in series to a battery. In this configuration, the plate connected to the battery's positive terminal develops a positive charge, while the plate attached to the negative terminal becomes negatively charged. An equal magnitude of charge is induced on the...
Trigonometric Fourier series
The trigonometric Fourier series specifically expresses a periodic function with a defined period T using sine...
Convergence of Fourier Series
The Gibbs phenomenon refers to the persistent oscillations and overshoots that occur near discontinuities...