Noise-robust bioimpedance approach for cardiac output measurement
Ethan K Murphy1, Justice Amoh1, Saaid H Arshad1
1Thayer School of Engineering, Dartmouth College, Hanover, NH 03755, United States of America.
Objective:
Congestive heart failure is a problem affecting millions of Americans. A continuous, non-invasive, telemonitoring device that can accurately monitor cardiac metrics could greatly help this population, reducing unnecessary hospitalizations and cost.
Approach:
Machine learning (ML) algorithms trained on electrical-impedance tomography (EIT) data are presented for portable cardiac monitoring. The approach was validated on a simulated thorax and a measured tank experiment. A highly detailed 4D chest model (finite element method mesh and conductivity profiles) was developed utilizing the 4D XCAT phantom to provide realistic data. The ML algorithms were trained using databases that assumed the presence of poorly contacting electrodes without any assumptions of knowing which electrodes would be bad in the experiment. The trained ML algorithms were compared to EIT evaluated with and without removing bad electrodes.
Main Results:
A regression support vector machine and a deep neural network (DNN) were found to be the most accurate and robust to poorly contacting electrodes while not needing to know which electrodes were in poor contact in the simulated and measured experiments, respectively.
Significance:
Although the ML algorithms are not always better than EIT (with bad electrodes removed), the comparable results without needing a priori knowledge of which electrodes are bad is seen as a very promising feature. An evaluation of computational costs showed that the DNN required comparable computational power to the other methods while requiring less memory, which could make the DNNs an attractive algorithm for a low-power, portable system. This work represents an important validation of the method using measured data, and model development, which is needed to apply this method on real clinical data. Additionally, the developed 4D simulated thorax model could be an important tool within the EIT community.
More Related Videos
07:43A Simple Approach to Perform TEER Measurements Using a Self-Made Volt-Amperemeter with Programmable Output Frequency
Published on: October 5, 2019
07:03Author Spotlight: Uncovering the Role of Mitochondrial Calcium Phosphate in Heart Failure and Bioenergetics
Published on: August 23, 2024
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Regulation of Water Output