Effective temporal resolution and image quality of volume scanning in 320-row detector CT
Atsushi Urikura1, Takanori Hara2, Tsukasa Yoshida3
1Division of Diagnostic Radiology, Shizuoka Cancer Center, 1007 Shimonagakubo, Nagaizumi, Sunto, Shizuoka, 411-8777, Japan. at.urikura@scchr.jp.
Abstract:
To measure the effective temporal resolution (eTR) and image quality for three reconstruction modes for non-helical volume scanning in area detector CT. Temporal sensitivity profiles (TSPs) were obtained and the full width of the TSP at half maximum was used as an index of the eTR. Image quality was assessed by image noise and the corrected artifact index. The half reconstruction mode had a higher eTR than the full and automatic patient motion collection (APMC) reconstructions. Compared to full reconstruction, the image noise with APMC and half reconstruction were increased by 16% and 35%. The corrected artifact index was lowest with APMC. The square root of full width at tenth maximum of the TSP showed a high coefficient of determination (R2 = 0.934) for image noise. This study revealed the TSPs and eTRs for non-helical volume scanning in area detector CT. A high eTR resulted in higher image noise.
Related Concept Videos
¹H NMR of Labile Protons: Temporal Resolution
The –OH proton in alcohols typically appears in the range of δ 2 to 5 ppm but can vary depending on the specific...
¹H NMR of Conformationally Flexible Molecules: Temporal Resolution
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Leaky Scanning
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT