Keratometry in children: Comparison between auto-refractokeratometer, rotating scheimpflug imaging, and biograph
Hassan Hashemi1, Samira Heydarian2, Mehdi Khabazkhoob3
1Noor Research Center for Ophthalmic Epidemiology, Noor Eye Hospital, Tehran, Iran.
Purpose:
To determine the agreement and validity of keratometric measurements in children with the Nidek ARK-510A auto-refractokeratometer compared to rotating Scheimpflug imaging with Pentacam and biograph with Lenstar LS 900.
Methods:
This study was conducted on 5620 schoolchildren aged 6-12 years in Shahroud, Iran. Minimum and maximum keratometry values and corneal astigmatism magnitude were compared by calculation of Paired difference, interclass correlation coefficient, and 95% limits of agreement (LoA) between devices.
Results:
After applying the exclusion criteria, 4215 right eyes were enrolled into the study. Mean minimum keratometry with Nidek ARK-510A, Pentacam, and Lenstar was 43.13±1.51, 43.14±1.48, and 42.87±1.46 diopters (D), respectively, and mean maximum keratometry was 43.97±1.59, 44.00±1.56, and 43.75±1.54D, respectively. Nidek ARK-510A overestimated minimum and maximum keratometry by 0.25±0.37 and 0.22±0.41, respectively, compared to Penatcam. The LoA between Nidek ARK-510A and Pentacam for minimum and maximum keratometry measurements were -0.98 to 0.47D and -1.02 to 0.57D, respectively. The LoA between Nidek ARK-510A and Lenstar for minimum and maximum keratometry measurements were -0.70 to 0.72D and -0.79 to 0.85D, respectively. The agreement between devices was best in emmetropes, worst in hyperopes. For astigmatic vector components, the agreements between devices were poor but best agreement was between Nidek ARK-510A and Pentacam.
Conclusions:
Keratometry measurement with Nidek ARK-510A was not significantly different from Pentacam and Lenstar, and this device can be used in screening programs in emmetropes.
More Related Videos
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Kinematic Equations for Rotation
For instance, imagine a point A on a rigid body engaged in circular motion. The translational velocity of this particular point can be calculated by taking the time derivatives of the displacement equation, which essentially measures the...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Rotation of Asymmetric Top
The relationship between the angular momentum of any rigid body and its angular velocity, both of which are vectors, involves the moment of inertia. The moment of inertia is a scalar quantity only for spherically symmetric...
Rotational Motion about a Fixed Axis
Rotation with Constant Angular Acceleration - I
Using our intuition, we can begin to see how rotational quantities such as angular displacement, angular velocity, angular acceleration, and time are related to one another. For example, if a flywheel...