Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography
R Ghanbarnezhad Farshi1,2, A Mesbahi2, M Johari3
1Immunology Research Center, Tabriz University of Medical Sciences, Tabriz, Iran.
Background:
While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free.
Objective:
The aim of this study was to evaluate the exposure doses received by patients during maxillofacial imaging with CBCT.
Methods:
Entrance surface dose (ESD) was measured by using thermoluminescent dosimeters (TLDs) attached to the eyes lids, parotid glands and thyroid of 64 patients in two imaging centers (A and B). Phantom dosimetry was performed by a cylindrical poly-methyl methacrylate (PMMA) head-size phantom and an ionization chamber for different exposure parameters. NewTom VGi and Planmeca Promax 3D CBCT scanners were used at centers A and B, respectively.
Results:
The mean ESD of the eyes, parotid glands and thyroid were 2.57, 2.33 and 0.28 mGy in center A, 0.35, 2.11 and 0.37 mGy in center B, respectively. ESD of the eyes revealed a significant difference in two centers; in center B, it was 86.4% lower than center A. In the phantom dosimetry, the measured doses of NewTom VGi were 2.63 and 2.08 mGy, respectively by changing field of view (FOV) size from 8×8 cm2 (height × diameter) to 6×6 cm2. For Planmeca Promax 3D, it ranged from 0.98 to 3.24 mGy depending on exposure parameters.
Conclusion:
There is a wide range of radiation doses dependent on the units, patients and selected scan parameters. Inappropriate selection of exposure settings, especially FOV size, can seriously increase patient dose.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Critical Region, Critical Values and Significance Level
In hypothesis testing, a sample statistic is converted to a test statistic using z, t, or chi-square distribution. A critical region is an area under the curve in probability distributions demarcated by the critical value. When the test statistic falls in this region, it suggests that the null hypothesis must be rejected. As this region contains all those values of the...
Critical Values
Critical Thinking
Colleges and universities are...