Comparison between 8- and 32-channel phased-array receive coils for in vivo hyperpolarized 13 C imaging of the human
Adam W Autry1, Jeremy W Gordon1, Lucas Carvajal1
1Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco.
Purpose:
To compare the performance of an 8-channel surface coil/clamshell transmitter and 32-channel head array coil/birdcage transmitter for hyperpolarized 13 C brain metabolic imaging.
Methods:
To determine the field homogeneity of the radiofrequency transmitters, B1 + mapping was performed on an ethylene glycol head phantom and evaluated by means of the double angle method. Using a 3D echo-planar imaging sequence, coil sensitivity and noise-only phantom data were acquired with the 8- and 32-channel receiver arrays, and compared against data from the birdcage in transceiver mode. Multislice frequency-specific 13 C dynamic echo-planar imaging was performed on a patient with a brain tumor for each hardware configuration following injection of hyperpolarized [1-13 C]pyruvate. Signal-to-noise ratio (SNR) was evaluated from pre-whitened phantom and temporally summed patient data after coil combination based on optimal weights.
Results:
The birdcage transmitter produced more uniform B1 + compared with the clamshell: 0.07 versus 0.12 (fractional error). Phantom experiments conducted with matched lateral housing separation demonstrated 8- versus 32-channel mean transceiver-normalized SNR performance: 0.91 versus 0.97 at the head center; 6.67 versus 2.08 on the sides; 0.66 versus 2.73 at the anterior; and 0.67 versus 3.17 on the posterior aspect. While the 8-channel receiver array showed SNR benefits along lateral aspects, the 32-channel array exhibited greater coverage and a more uniform coil-combined profile. Temporally summed, parameter-normalized patient data showed SNRmean,slice ratios (8-channel/32-channel) ranging 0.5-2.00 from apical to central brain. White matter lactate-to-pyruvate ratios were conserved across hardware: 0.45 ± 0.12 (8-channel) versus 0.43 ± 0.14 (32-channel).
Conclusion:
The 8- and 32-channel hardware configurations each have advantages in particular brain anatomy.
More Related Videos
11:57Measuring the Spin-Lattice Relaxation Magnetic Field Dependence of Hyperpolarized [1-13C]pyruvate
Published on: September 13, 2019
14:56Investigating Cardiac Metabolism in the Isolated Perfused Mouse Heart with Hyperpolarized [1-13C]Pyruvate and 13C/31P NMR Spectroscopy
Published on: April 21, 2023
Related Concept Videos
Receiver Operating Characteristic Plot
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Carbon-13 (¹³C) NMR: Overview
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Phase Diagrams