Imaging Corrosion at the Metal-Paint Interface Using Time-of-Flight Secondary Ion Mass Spectrometry
Published on: May 6, 2019
Improved gating device of time-of-flight ion mass analyzer for ion sources
Kyumin Choe1, DongHwan Kim1, Kyoung-Jae Chung1
1Department of Nuclear Engineering, Seoul National University, Seoul 08826, South Korea.
Abstract:
A gating device is a critical component in a time-of-flight (TOF) ion mass analyzer for identifying ion species extracted from ion sources according to their mass-to-charge ratio. It consists of several concentric ring electrodes to deflect as many ions as possible within a very short gate time, thus increasing the ion current collected in a Faraday cup. In this study, we further improve the ion collection efficiency of the gating device by minimizing potential distortion along the paths of ions, by modifying the structure of gate electrodes and applying a bipolar gate pulse on them. Using the improved gating device, we find that the ion beam current increased by nearly 50% compared to the conventional one without changing the noise level. The improved TOF system is well applied to identify the spectrum of ion species extracted from a cold cathode Penning ion source, depending on the operating conditions.
More Related Videos
05:52Author Spotlight: Enhanced Histone PTM Isomer Identification Through LC-TIMS-ToF MS/MS and PASEF
Published on: January 12, 2024
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Non-gated Ion Channels
Compared to the gated ion channels, the non-gated channels, also known as leakage or passive channels, have no gating mechanism....
Ligand-gated Ion Channels
Three Subfamilies of Ligand-gated Ion Channels
Ligand-gated ion channels fall into three subfamilies. The 'Cys-loop' includes the nicotinic acetylcholine receptors, γ-aminobutyric acid (GABA), glycine, and 5-hydroxytryptamine receptors. The second one is the 'Pore-loop' channels that...
Voltage-gated Ion Channels
Generally, all voltage-gated ion channels have a 'voltage-sensing domain' that spans the lipid bilayer. The charged residues in the sensor move in response to the membrane potential changes that open the channel allowing ions movement. There are several types of...
Mechanically-gated Ion Channels
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...
G-Protein Gated Ion Channels
Sensory...