The reliability of cutaneous low-frequency oscillations in young healthy males
Gary J Hodges1, Matthew M Mallette1, Stephen S Cheung1
1Environmental Ergonomics Laboratory, Department of Kinesiology, Brock University, St. Catharines, Ontario, Canada.
Objective:
Spectral analyses of laser-Doppler flowmetry measures enable a simple and non-invasive method to investigate mechanisms regulating skin blood flow. We assessed within-day and day-to-day variability of cutaneous spectral analyses.
Methods:
Eleven young, healthy males were tested twice in three identical sessions, with 19 to 24 days between visits, for a total of six tests. Wavelet data were analyzed at rest, in response to local skin heating to 42 and 44°C, and during 5-minutes PORH. We did this for six frequency bands commonly associated with physiological functions. To assess reliability, we calculated CV and ICC scores.
Results:
At rest, mean CV for the wavelet data ranged from 21% to 24% and ICC scores ranged from 0.67 to 0.91. During local heating, mean CV scores ranged from 17% to 22% and mean ICC scores ranged from 0.71 to 0.95. For peak PORH, CV ranged from 14% to 23% and the ICC range was 0.88 to 0.97. For the area under the curve of the PORH, CV range was 12% to 21% and ICC range was 0.81 to 0.92.
Conclusions:
These analyses indicate good-to-excellent reliability of the wavelet data in healthy young males.