Cyclic and Load-to-Failure Properties of All-Suture Anchors in Human Cadaveric Shoulder Glenoid Bone
John A Ruder1, Ephraim Y Dickinson2, Richard D Peindl3
1Department of Orthopaedic Surgery, Carolinas Medical Center, Charlotte, North Carolina, U.S.A.
Purpose:
To evaluate the cyclic displacement and ultimate load to failure of 4 all-suture anchors in human cadaveric shoulder glenoid bone.
Methods:
Four all-suture anchors indicated for glenoid labral repair were tested in 14 matched pairs of human cadaveric fresh-frozen glenoids. Anchors were inserted at 4 different locations for a total of 112 tests (12-, 3-, 6-, and 9-o'clock positions for right glenoids). Cyclic loading (10 to 60 N at 1 Hz for 200 cycles) and single pull-to-failure testing (33 mm/s) were performed. A Kruskal-Wallis 1-way analysis of variance with the Dunn multiple-comparison post hoc test was used for statistical analysis.
Results:
One matched pair was excluded because of poor bone quality. Thus, 13 matched pairs were included in the study, and a total of 104 tests were performed. The Q-Fix anchors showed significantly less displacement after 100 cycles (mean ± standard deviation, 1.40 ± 0.97 mm; P < .001) and 200 cycles (1.53 ± 1.00 mm, P < .001) than all other anchors tested. The Q-Fix (191.3 ± 65.8 N), Suturefix (188.3 ± 61.4 N), and JuggerKnot (183.6 ± 63.5 N) anchors had significantly greater ultimate loads to failure than the Iconix anchors (143.5 ± 54.1 N) (P = .01, P = .012, and P = .021, respectively). Displacement greater than 5 mm occurred in 6 Iconix anchors (22.1%), 5 Suturefix anchors (19.2%), 4 JuggerKnot anchors (15.4%), and 0 Q-Fix anchors (0%).
Conclusions:
The Q-Fix anchors showed less displacement with cyclic loading than the Iconix, JuggerKnot, and Suturefix anchors. The Iconix anchors had a lower ultimate load to failure than the Q-Fix, Suturefix, and JuggerKnot anchors. Only the Q-Fix group had no anchors displace greater than 5 mm with cyclic loading.
Clinical Relevance:
All-suture anchors vary in their deployment mechanism, which may alter their strength and performance. Operators must be aware of these anchors' propensity to displace while deploying them.
Related Concept Videos
The Anchoring-and-Adjustment Heuristic
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
Lipids as Anchors
The carboxy-terminal of most of the prenylated proteins, such as Ras proteins, contains...
Anchoring Junctions
GPI Anchoring of Proteins in the ER Membrane
GPI-anchor structure
A sequence of 11 enzymatic reactions results in the synthesis of the complete GPI anchor consisting of a hydrophobic and a hydrophilic portion. The hydrophobic portion comprises phosphatidylinositol, while the hydrophilic part comprises polar groups like phosphoethanolamine,...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...