Related Experiment Video
Updated: Jan 26, 2026

Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Early post-Humanitarian Device Exemption experience with the Neuroform Atlas stent
Jenny P Tsai1, Julian Hardman1, Nina Z Moore1
1Cerebrovascular Center, Cleveland Clinic, Cleveland, Ohio, USA.
Introduction:
The low-profile Neuroform Atlas stent received FDA Humanitarian Device Exemption status (HDE) in January 2018 for stent-assisted coil embolization of wide-necked saccular aneurysms. We review and report our results with the Atlas stent in our institution within the first year after its HDE approval.
Methods:
Our retrospective chart review identified patients treated with the Atlas stent. We analyzed the patient demographics, aneurysm characteristics, stent parameters and configuration, complications, angiographic, and clinical outcomes at discharge.
Results:
From January to December 2018, 76 Atlas stents were deployed in 58 patients (average 1.3 stents/patient). Median patient age was 63.5 (IQR 56-71) years. Fifty-six (96.6%) patients had elective embolization of unruptured aneurysms, while two (3.4%) patients underwent embolization of a ruptured aneurysm within 2 weeks of subarachnoid hemorrhage. Forty (69.0%) patients were treated with a single stent, 15 (25.9%) with a Y-stent, and three (5.2%) with X-stent configuration. All stent deployments were technically successful. Most stents (82.9%) were the smallest 3 mm diameter devices. Procedural complications included transient stent-associated thrombosis in three (5.2%) patients and aneurysm rupture in one (1.7%). None had distal embolization, associated cerebral infarction, or permanent neurological deficits. Immediate Raymond-Roy 1 occlusion was achieved in 41 (70.7%) patients. Median hospital length of stay for elective aneurysm embolization was 1 day. Excellent outcomes with median National Institute of Health Stroke Scale score 0 (IQR 0-0) and modified Rankin Score 0 (IQR 0-1) were seen for elective patients at discharge.
Conclusion:
The Neuroform Atlas stent provided a reliable technical and safety profile for the treatment of intracranial wide-neck aneurysms. Further experience is needed to determine long-term durability and safety of this device.
Related Concept Videos
What is an Experiment?
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...
Controls in Experiments
Randomized Experiments
Simple randomization
Simple...
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
The Stanford Prison Experiment