Strategies for radiation exposure-sparing in fluoroscopically guided invasive cardiovascular procedures
Paul N Fiorilli1,2,3, Taisei Kobayashi1,2,3, Jay Giri1,2,4,3
1Cardiovascular Medicine Division, Hospital of the University of Pennsylvania, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania.
Background:
Minimizing radiation exposure during x-ray fluoroscopically guided procedures is critical to patients and to medical personnel. Tableside adjustment of x-ray image acquisition parameters can vary the fluoroscopic radiation exposure rate.
Objectives:
To determine the impact of adjusting four tableside controllable image acquisition parameters on x-ray fluoroscopic radiation exposure rate.
Methods:
We made fluoroscopic exposures of a standard radiologic phantom to measure radiation exposure rates as kerma•area product per second of exposure and milligray per x-ray pulse under all possible combinations of detector zoom mode, collimated image field size, fluoroscopy dose mode, and fluoroscopy pulse frequency.
Results:
Kerma•area product per second was linearly proportional to pulse frequency. Selecting larger detector zoom modes, smaller collimated image field sizes and low dose fluoroscopy mode each decreased exposure rate. We found a > 20-fold variation in dose rates over the range of acquisition parameter combinations.
Conclusions:
Selecting the most appropriate fluoroscopy acquisition parameters enables physician operators to adjust radiation exposure rates over a large range. Judicious selection of acquisition parameters can reduce patient and medical personnel radiation exposure by as much as 95% compared to "standard" fluoroscopy protocol settings.
Related Concept Videos
Biological Effects of Radiation
Spare Receptors
Radiation: Applications
The average...
Overview of the Cardiovascular System
Heart
The heart is the central pump of the cardiovascular system that circulates blood throughout the body. It comprises two atria receiving the blood and two ventricles pumping blood out of the heart. Their rhythmic contractions, called heartbeats, ensure that blood flow remains continuous.
Blood Vessels
Blood...
Absorption of Radiation
Generating Electromagnetic Radiations