Dosimetric comparison of short and full arc in spinal PTV in volumetric-modulated arc therapy-based craniospinal
Biplab Sarkar1, Anusheel Munshi1, Tharmarnadar Ganesh1
1Radiation Oncology, Manipal Hospitals, New Delhi, Delhi 110075, India.
Abstract:
Since 2011 when it was first described, the volumetric-modulated arc therapy (VMAT) technique for craniospinal irradiation (CSI) has always seen the use of large arc lengths for the spine fields ranging from 200° to 360°. This study was aimed to do a dosimetric comparison between the large and shorter spinal arc for CSI. For a cohort of 10 patients, 2 VMAT CSI plans were created for each patient, one using the conventional full 360° arc (VMAT_FA) for the spine and the other using 100° posterior arc (VMAT_PA) for 23.4 Gy and 35 Gy prescriptions. In both the plans, 360° arc fields were employed for treating cranial volume. Spillage dose (DBody-PTV) to Body-PTV (DBody-PTV: dose to body excluding planning target volume) was compared with VMAT_FA and VMAT_PA plans. In addition to these VMAT plans, a 3-dimensional conformal radiotherapy plan was also created for all these patients to compare the DBody-PTV and target volume related dose constraints. Mean D95% difference between the two VMAT plans did not exceed 1.3% for cranial and spinal targets for both prescription levels. The conformity index (CI) was averaged over both prescription doses. Average CI shows a similar value for VMAT_FA (0.84 ± 0.04) and VMAT_PA (0.82 ± 0.05) plans. D95%, V110% and CI did not exhibit a statistically significant difference between partial and full-arc VMAT plans. However, the VMAT_PA plan exhibited a lower DBody-PTV compared to VMAT_FA plans (0.007 ≤ p < 0.05) in the 1 to 5 Gy range. Nevertheless, partial arc plans could not offer a statistically significant dose reduction for delineated organs compared to full arc plans, except for bilateral kidneys.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Gene Therapy
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits