Calibration of blood pressure measurements in the Jackson Heart Study
Samantha R Seals1,2, Lisandro D Colantonio3, Jonathan V Tingle4
1Department of Data Science, Center of Biostatistics and Bioinformatics.
Background:
In longitudinal research studies with follow-up examinations, the devices used to measure phenotypes may change over time. When a device change occurs, the two devices should be calibrated to each other to ensure that measurements are comparable. This paper details the Jackson Heart Study (JHS) blood pressure (BP) comparability study.
Participants And Methods:
During its second clinic exam (2005-2008), the JHS switched from a random-zero sphygmomanometer (RZS) BP measurement device to an oscillometric device (OD). During this exam, BP measurements from both an RZS and an OD were taken simultaneously in 2117 participants for the purpose of calibration. Five methods for calibrating systolic BP (SBP) and diastolic BP (DBP) were considered: ignoring the change, ordinary least squares regression, adding the average difference, Deming regression, and robust regression.
Results:
Using the RZS and OD, the mean (SD) SBP was 125.5 (19.2) and 126.5 (19.9), respectively, and the mean (SD) DBP was 76.4 (10.6) and 74.0 (11.0), respectively. The correlation between RZS and the OD was 0.90 for SBP and 0.80 for DBP. The prevalence of high BP and hypertension and associations with albuminuria were similar when applying each of the five calibration methods. Robust regression was chosen for calibration, giving the following equations:(Equation is included in full-text article.)These equations had a higher R statistic than using calibration equations from the Coronary Artery Risk Development in Young Adults Study and the Heinz Nixdorf Recall Study.
Conclusions:
The JHS BP data have been calibrated using the above equations for use in future analyses.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...