Comparison of Salbutamol Delivery Efficiency for Jet versus Mesh Nebulizer Using Mice
Kyung Hwa Chang1, Sang-Hyub Moon2, Jin Young Oh3
1Department of Medical Engineering, Dongguk University College of Medicine, Goyang-si, Gyeonggi-do 10326, Korea. chang9429@hotmail.com.
Abstract:
Recent reports using a breathing simulator system have suggested that mesh nebulizers provide more effective medication delivery than jet nebulizers. In this study, the performances of jet and mesh nebulizers were evaluated by comparing their aerosol drug delivery efficiencies in mice. We compared four home nebulizers: two jet nebulizers (PARI BOY SX with red and blue nozzles), a static mesh nebulizer (NE-U22), and a vibrating mesh nebulizer (NE-SM1). After mice were exposed to salbutamol aerosol, the levels of salbutamol in serum and lung were estimated by ELISA. The residual volume of salbutamol was the largest at 34.6% in PARI BOY SX, while the values for NE-U22 and NE-SM1 mesh nebulizers were each less than 1%. The salbutamol delivery efficiencies of NE-U22 and NE-SM1 were higher than that of PARI BOY SX, as the total delivered amounts of lung and serum were 39.9% and 141.7% as compared to PARI BOY SX, respectively. The delivery efficiency of the mesh nebulizer was better than that of the jet nebulizer. Although the jet nebulizer can generate smaller aerosol particles than the mesh nebulizer used in this study, the output rate of the jet nebulizer is low, resulting in lower salbutamol delivery efficiency. Therefore, clinical validation of the drug delivery efficiency according to nebulizer type is necessary to avoid overdose and reduced drug wastage.
Related Concept Videos
Free Jet
Mesh Analysis
A fundamental concept in mesh analysis is the definition of meshes and mesh currents. A mesh is a closed...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Mesh Analysis with Current Sources
Current Source in One Mesh: The analysis process is straightforward when a current source is found in only one mesh within the circuit. Mesh currents are assigned as usual, with the mesh containing the current source excluded from the analysis. Kirchhoff's voltage law...
Mesh Analysis for AC Circuits
The process of harmonizing these impedances begins with a clear understanding of the input and output signals. Once these signals are known, the...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...