Related Experiment Video
Updated: Jan 25, 2026

Atom Probe Tomography Analysis of Exsolved Mineral Phases
Published on: October 25, 2019
Assessing the Spatial Accuracy of the Reconstruction in Atom Probe Tomography and a New Calibratable Adaptive
Anna V Ceguerra1, Alec C Day1, Simon P Ringer1
1Australian Centre for Microscopy & Microanalysis, and School of Aerospace, Mechanical and Mechatronic Engineering,The University of Sydney,Sydney, NSW 2006,Australia.
Abstract:
We define a measure for the accuracy of tomographic reconstruction in atom probe tomography, named here the spatial error index. We demonstrate that this index can be used to compare rigorously the spatial accuracy of various different approaches to the calculation of tomographic reconstruction. This is useful, for example, to evaluate the performance of alternate tomographic reconstruction approaches, and ensures that the comparisons are independent of individual data quality or other instrumental parameters. We then introduce a new "adaptive reconstruction" formalism that uses a progression of reconstruction parameters based on a per-atom correction from the cube root of the inverse of the voltage, along with linear correction factors linked to the evaporation sequence. We apply the measure for spatial accuracy to this new reconstruction protocol.
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Reconstruction of Signal using Interpolation
Improving Translational Accuracy
Atomic Structure
Atomic Mass
Uncertainty in Measurement: Accuracy and Precision