Absolute radiotracer concentration measurement using whole-body solid-state SPECT/CT technology: in vivo/in vitro
John A Kennedy1,2, Ilya Reizberg3, Rachel Lugassi3
1Department of Nuclear Medicine, Rambam Health Care Campus, P.O.B. 9602, 3109601, Haifa, Israel. j_kennedy@rambam.health.gov.il.
Solid-state SPECT/CT imaging accurately quantifies radiotracer levels in the bladder. This non-invasive method, with less than 20% error, aids in diagnosis and treatment response assessment.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Quantitative Analysis
Background:
- Quantitative clinical software is crucial for accurate medical imaging analysis.
- Solid-state Cadmium-Zinc-Telluride (CdZnTe) SPECT/CT cameras offer advanced imaging capabilities.
- Validating software accuracy with in vivo and in vitro measurements is essential.
Purpose of the Study:
- To determine the accuracy of recently approved quantitative clinical software for solid-state SPECT/CT.
- To compare in vivo measurements from SPECT/CT with in vitro measurements from urine samples.
- To assess the reliability of quantitative software in measuring radiotracer concentration in the urinary bladder.
Main Methods:
- Performed Bone SPECT/CT on 16 patients using technetium-99m methylene diphosphonic acid.
- Collected urine samples for in vitro radiopharmaceutical concentration measurement.
- Used quantitative software (Q.Metrix, GE Healthcare) to measure in vivo radiotracer concentration and Standardized Uptake Value (SUV) in the bladder from SPECT/CT images.
- Employed linear regression to validate in vivo/in vitro relations.
Main Results:
- Thirteen in vivo/in vitro pairs were analyzed after excluding outliers.
- All linear regressions showed strong identity relations (R² ≥ 0.85, P < 0.0001).
- Average SUV measurements across users yielded a slope of 0.99 ± 0.25 and intercept of 0.14 ± 5.15 g/mL, with an average residual difference < 20%.
Conclusions:
- Solid-state SPECT/CT imaging reliably quantifies in vivo urinary bladder radiotracer concentrations with approximately 20% accuracy.
- The quantitative software demonstrates accuracy in measuring radiotracer uptake.
- This non-invasive in vivo quantification method holds potential for improved diagnosis and treatment response assessment.
Related Concept Videos
Drug Concentrations: Measurements
Plasma...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Concentration Cells
Consider the following voltaic cell:
Reliability and Validity
Mean Absolute Deviation
Let us consider a dataset containing the number of unsold cupcakes in five shops: 10, 15, 8, 7, and 10. Initially, calculate the sample mean. Then calculate the deviation, or the difference, between each data value and the mean. Next, the absolute values of these deviations are added and divided by the sample size to...
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....