Determination of amphetamine-type stimulants in urine samples using microextraction by packed sorbent and gas
Sara Malaca1, Tiago Rosado1, José Restolho2
1Centro de Investigação em Ciências da Saúde, Universidade da Beira Interior (CICS-UBI), Covilhã, Portugal; Laboratório de Fármaco-Toxicologia - UBIMedical, Universidade da Beira Interior, Covilhã, Portugal.
Abstract:
The aim of this work was the development, optimization and full validation of a method applying microextraction by packed sorbent (MEPS) coupled to gas chromatography-mass spectrometry (GC-MS) to determine amphetamine (AMP), methamphetamine (MAMP), 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxyethylmethamphetamine (MDMA), 3,4-methylenedioxy-N-methyl-α-ethylfenilethylamine (MBDB), and 3,4-methylenedioxy-N-ethylamphetamine (MDE) in urine samples. Using 200 μL of sample, the MEPS procedure was optimized concerning type of sorbent, sample dilution, number of strokes, activation of the ion exchange mechanism and composition of both washing and elution solvents. The method was fully validated according to the Food and Drug Administration and the Scientific Working Group of Forensic Toxicology guidelines for the validation of bioanalytical methods. The studied parameters included selectivity, calibration model and linearity, limits of detection and quantification, precision, accuracy, stability, dilution integrity and recoveries. Linearity was obtained in the range of 25-1000 ng/mL for MAMP, MBDB and MDE, 35-1000 ng/mL for AMP and MDMA, and 50-1000 ng/mL for MDA, with coefficients of determination (R2) >0.99 for all analytes. Both intra- and inter-day precision and accuracy were adequate, and coefficients of variation lower than 15% and mean relative errors (RE) within a range of ±15% of the theoretical concentrations were obtained for all compounds under study. Analyte recoveries ranged from 19 to 71%, allowing LLOQs ≤50 ng/mL.
More Related Videos
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
05:29Profiling Volatile Compounds in Blackcurrant Fruit using Headspace Solid-Phase Microextraction Coupled to Gas Chromatography-Mass Spectrometry
Published on: June 9, 2021
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
CNS Stimulants: Cocaine, Amphetamines and Cannabinoids