The Importance of the Location of Dorsal Root Ganglion Stimulator Electrodes Within the Nerve Root Exit Foramen
Related Experiment Video
Updated: Jan 25, 2026

Analysis of Immune Cells in Single Sciatic Nerves and Dorsal Root Ganglion from a Single Mouse Using Flow Cytometry
Published on: December 6, 2017
Sean Martin1, George Hadjipavlou1,2, Rodrigo Garcia Ortega3
1Nuffield Department of Surgical Sciences, University of Oxford, Oxford, UK.
Objective:
To quantify the relationship between the electrical power requirement to achieve pain relief and the position of the active electrode of dorsal root ganglion stimulators within the spinal nerve root exit foramen.
Materials And Methods:
Retrospective analysis of prospectively collected data of 92 consecutive patients undergoing dorsal root ganglion stimulation (DRGS) for chronic pain in a single center. Cervical and sacral cases, and failed trials/explanted DRGS were excluded, so we report on 57 patients with 78 implanted leads. Anteroposterior and lateral fluoroscopic images of the lead in the exit foramen were examined, and the active electrode positions were put into categories depending on their location relative to fixed anatomical landmarks. The clinical outcome and the power requirements for each of these groups of electrodes were then analyzed. Overall pain outcome was assessed by numeric pain rating scale score pre-operatively and post-operatively.
Results:
There was no significant relationship between power requirements and mediolateral electrode position, although the lowest average was observed with electrode positions directly below the center of the pedicle. On lateral x-ray, the lowest power requirements were observed in the electrodes positioned superodorsally or dorsally within the foramen. Importantly, power requirements in this location were consistently low, while the power requirements in other locations were not only higher but also much more variable. Electrodes in the superodorsal position required a median output power almost four times lower than electrodes in other positions (p = 0.002). Clinical outcome was not significantly related to power requirement or foraminal position.
Conclusion:
Aiming for a superodorsal electrode position on lateral intraoperative fluoroscopy is desirable, since siting leads in this location reduces the required stimulator output power very substantially and thus will extend battery life. Position within the foramen does not determine clinical outcome, and so the implanter can safely aim for the low power site without detriment to the analgesic efficacy of the system.
Related Concept Videos
Primary and Secondary Growth in Roots and Shoots
Root Mean Square
For example, consider the velocity of gas molecules in a container. The gas molecules are moving in different directions, which might impart positive and negative...
Root-Locus Method
This system can be represented by a block...
Construction of Root Locus
For positive gain values, the root locus exists on the real axis to the left of an odd number of finite open-loop poles or zeros. The root locus starts at the open-loop poles and traces the paths of the closed-loop poles as the gain...
Properties of the Root Locus
To determine if a point lies on the root locus, the criterion involves the sum of angles contributed by all poles and zeros to that point. Specifically, this sum must be an odd multiple of 180 degrees. The gain at any point on...
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...