Development of Ultrafast Separations Using Negative Pulse Partial Modulation To Enable New Directions in Gas
Derrick V Gough1, Dong H Song1, Sonia Schöneich1
1Department of Chemistry , University of Washington , Box 351700, Seattle , Washington 98195 , United States.
Abstract:
Partial modulation via a pulse flow valve operated in the negative pulse mode is developed for high-speed one-dimensional gas chromatography (1D-GC), comprehensive two-dimensional (2D) gas chromatography (GC × GC), and comprehensive three-dimensional gas chromatography (GC3). The pulse flow valve readily provides very short modulation periods, PM, demonstrated herein at 100, 200, and 300 ms, and holds significant promise to increase the scope and applicability of GC instrumentation. The negative pulse mode creates an extremely narrow, local analyte concentration pulse. The reproducibility of the negative pulse mode is validated in a 1D-GC mode, where a pseudosteady-state analyte stream is modulated, and 8 analytes are baseline resolved (resolution, Rs ≥ 1.5) in a 200 ms window, providing a peak capacity, nc, of 14 at unit resolution ( Rs = 1.0). Additionally, the pulse width, pw, of the pulse flow valve "injection" relationship to peak width-at-base, wb, resolution between peaks and detection sensitivity are studied. To demonstrate the applicability to GC × GC, a high-speed separation of a 20-component test mixture of similar, volatile analytes is shown. Analytes were separated on the second-dimension column, 2D, with 2 wb ranging from 7 to 12 ms, providing an exceptional 2D peak capacity, 2 nc, of ∼12 using a modulation period ( PM) of 100 ms. Next, a 12 min separation of a diesel sample using a PM of 300 ms is presented. The 1 wb is ∼4 s, resulting in a 1 nc of ∼180, and 2 wb is ∼18 ms, resulting in a 2 nc of ∼17, thus achieving a nc,2D of ∼3000 in this rapid GC × GC diesel separation. Finally, GC3 with time-of-flight mass spectrometry (TOFMS) detection using a PM of 100 ms applied between the 2D and 3D columns is reported. Narrow third dimension, 3D, peaks with 3 wb of ∼15 ms were obtained, resulting in a GC3 peak capacity, nc,3D, of ∼35 000 in a 45 min separation.
More Related Videos
08:22Nitrogen Compound Characterization in Fuels by Multidimensional Gas Chromatography
Published on: May 15, 2020
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....