Falsely Elevated Vancomycin Concentrations in a Patient Not Receiving Vancomycin
Vivian Tsoi1, Vipin Bhayana2,3, Anne Marie Bombassaro2,3
1St. Michael's Hospital, Toronto, Ontario, Canada.
Abstract:
Therapeutic drug monitoring (TDM) of vancomycin is commonly performed using immunoassays. This case describes falsely elevated vancomycin serum concentrations, possibly secondary to endogenous protein interference. Vancomycin was prescribed for a patient with a suspected septic knee. A blood sample for TDM was inadvertently collected before the first dose. The reported concentration was 36.1 mg/L using the Roche Modular P analyzer and remained high over the next 48 hours and 8 months later in the absence of vancomycin therapy. Vancomycin was undetectable in the patient sample by liquid chromatography-tandem mass spectrometry. The sample was subsequently investigated for endogenous protein interference. The responsible interference was removed by polyethylene glycol precipitation and heat inactivation. Four alternative immunoassays with varying test principles measured vancomycin concentrations ranging from undetectable to 108 mg/L. A glucose-6-phosphate dehydrogenase detection method was common to the two immunoassays exhibiting the greatest interference. To our knowledge, this is the first report of falsely elevated vancomycin concentrations on the Roche Modular P analyzer. Immunoassays are generally robust in facilitating TDM but are susceptible to cross-reactivity. Assay interference should be considered and laboratory professionals contacted when vancomycin levels do not correlate with clinical expectations.
More Related Videos
09:14Surface Electromyographic Biofeedback as a Rehabilitation Tool for Patients with Global Brachial Plexus Injury Receiving Bionic Reconstruction
Published on: September 28, 2019
07:36An Experimental Analysis of Children's Ability to Provide a False Report about a Crime
Published on: May 3, 2016
Related Concept Videos
Freezing Point Depression and Boiling Point Elevation
The boiling point of a liquid is the temperature at which its vapor pressure is equal to ambient atmospheric pressure. Since the vapor pressure of a solution is lowered due to the presence of nonvolatile solutes, it stands to reason that the solution’s boiling point will subsequently be increased. Vapor pressure increases with temperature, and so a solution will require a higher temperature than will pure solvent to achieve any given vapor pressure, including one...
False Memories
One primary source of false memories is misattribution, where individuals incorrectly associate external information...
Receiver Operating Characteristic Plot
Elevation of Intermediate Points on Vertical Curves
Concentration Cells
Consider the following voltaic cell:
Solution Concentration and Dilution