Noninvasive Determination of Blood Pressure by Heart Sound Analysis Compared With Intra-Arterial Monitoring in
Gaurav Kapur1, Lingguang Chen2, Yong Xu3
1Department of Pediatrics, Children's Hospital of Michigan/Wayne State University, Detroit, MI.
Objectives:
To develop a novel device to predict systolic and diastolic blood pressure based on measured heart sound signals and evaluate its accuracy in comparison to intra-arterial blood pressure readings.
Study Design:
Prospective, observational pilot study.
Setting:
PICU.
Patients:
Critically ill children (0-18 yr) undergoing continuous blood pressure monitoring via radial artery intra-arterial catheters were enrolled in the study after informed consent. The study included medical, cardiac, and surgical PICU patients.
Interventions:
Along with intra-arterial blood pressure, patient's heart sounds were recorded simultaneously by a highly sensitive sensor taped to the chest. Additional hardware included a data acquisition unit and laptop computer. Subsequently, advanced signal processing technologies were used to minimize random interfering signals and extract and separate S1 and S2 signals. A computerized model was then developed using artificial neural network systems to estimate blood pressure from the extracted heart sound analysis.
Measurements And Main Outcomes:
We found a statistically significant correlation for systolic (r = 0.964; R = 0.928) and diastolic (r = 0.935; R = 0.868) blood pressure readings (n = 491) estimated by the novel heart-sound signal-based method and those recorded by intra-arterial catheters. The mean difference of the individually paired determinations of the blood pressure between the heart-sound-based method and intra-arterial catheters was 0.6 ± 7 mm Hg for systolic blood pressure and -0.06 ± 5 mm Hg for diastolic blood pressure, which was within the recommended range of 5 ± 8 mm Hg for any new blood pressure devices.
Conclusions:
Our findings provide proof of concept that the heart-sound signal-based method can provide accurate, noninvasive blood pressure monitoring.
Related Concept Videos
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Sound as Pressure Waves
The pressure fluctuation depends on the difference in displacements between the successive points in the...
Errors occurring during blood pressure monitoring
Several factors...
Intensity and Pressure of Sound Waves
Unlike the time average of a sinusoidal term, which is zero since it is positive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)