A rationale for cone beam CT with extended longitudinal field-of-view in image guided adaptive radiotherapy
K Mohamathu Rafic1, S Balasingh Timothy Peace1, Mathew Manu1
1Department of Radiation Oncology, Christian Medical College, Vellore 632004, Tamil Nadu, India.
Purpose:
To investigate the efficacy of using cone beam CT with extended longitudinal field-of-view (CBCTeLFOV) for image guided adaptive radiotherapy (IGART).
Methods:
The protocol acquires two CBCT scans with a linear translation of treatment couch in the patient plane, allowing a 1 cm penumbral overlap (i.e. cone beam abutment) and fused as a single DICOM set (CBCTeLFOV) using a custom-developed software script (coded in MatLab®) for extended localization. Systemic validation was performed to evaluate the geometric and Hounsfield Units accuracy at the overlapping regions of the CBCTeLFOV using a Catphan®-504 phantom. Two case studies were used to illustrate the CBCTeLFOV-based IGART workflow in terms of dosimetric and clinical perspectives. Segmentation accuracy/association between repeat CT (re-CT) and CBCTeLFOV was evaluated. Moreover, the efficacy of the CBCTeLFOV image data in deformable registration was also described.
Results:
Slice geometry, spatial resolution, line profiles and HU accuracy in the overlapping regions of the CBCTeLFOV yielded identical results when compared with reference CBCT. In patient studies, the dice-similarity-coefficient evaluation showed a good association (>0.9) between re-CT and CBCTeLFOV. Dosimetric analysis of the CBCTeLFOV-based adaptive re-plans showed excellent agreement with re-CT based re-plans. Moreover, a similar and consistent pattern of results was also observed using deformed image data (initial planning CT deformed to CBCTeLFOV) with extended longitudinal projection and the same frame-of-reference as that of the CBCTeLFOV.
Conclusion:
Utilization of CBCTeLFOV proves to be clinically appropriate and enables accurate prediction of geometric and dosimetric consequences within the planned course of treatment. The ability to compute CBCTeLFOV-based treatment plans equivalent to re-CT promises a potential improvement in IGART practice.
More Related Videos
Related Concept Videos
Longitudinal Research
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Longitudinal Studies