A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
Comparison between navigated reported position and postoperative computed tomography to evaluate accuracy in a
Francisco Figueroa1, Edgar Wakelin2, Joshua Twiggs2
1Clínica Alemana-Universidad del Desarrollo, Chile; Hospital Sótero del Río, Chile.
Background:
Computer navigation increases reproducibility compared to non-navigated total knee arthroplasty (TKA). Robotics navigation is a branch of computer navigation technology that might further improve accuracy of implant placement. The aim of this study is to assess the accuracy achieved in TKA with a robotic navigation system.
Methods:
One hundred seventy three knees. System studied: Omni navigation System (OMNI, Raynham, MA). Navigated femoral and tibial cuts were compared to postoperative computed tomography (CT). Measurements reviewed: femoral coronal alignment (FCA), femoral sagittal alignment (FSA), femoral rotational alignment (FRA), tibial coronal alignment (TCA), tibial sagittal alignment (TSA) and hip-knee-ankle (HKA) angle. Statistical analysis was made using R.
Results:
The mean differences between the navigated reported and the CT positions were: FCA: 0.1 ± 1.2° more varus (P = 0.58), FSA: 1.5 ± 0.3° more flexed (P < 0.001), FRA: 0.0 ± 1.7° (P = 0.93), TCA: 0.7 ± 1.1° more varus (P < 0.001), TSA: -1.3 ± 1.5 more negative slope (P < 0.001), HKA angle: 0.4 ± 2.4 more varus (P < 0.049). The percentages of concordance inside a three degree difference were: FCA: 98% (169 knees), FSA: 100% (173 knees), FRA: 94% (162 knees), TCA: 99% (171 knees), TSA: 93% (161 knees) and HKA angle: 83% (144 knees).
Conclusions:
The current study showed that the robotic navigation system studied is highly accurate regarding final implant positioning for FCA, FRA and TCA. It has less accuracy in FSA, TSA and the HKA angle.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Uncertainty in Measurement: Accuracy and Precision
Improving Translational Accuracy
Improving Translational Accuracy