The reversible adjustable coupling: A lightweight and low-cost alignment component for the lower limb prosthesis
Kazuhiko Sasaki1, Jutamat Pinitlertsakun1, Pakwan Nualnim1
1Sirindhorn School of Prosthetics and Orthotics, Faculty of Medicine, Mahidol University, Siriraj Hospital, Bangkok, Thailand.
A new, low-cost reversible adjustable coupling for lower limb prosthetics was developed. This lightweight component offers essential alignment adjustments, making prosthetic care more accessible in low-income countries.
Area of Science:
- Biomedical Engineering
- Prosthetics and Orthotics
Background:
- Lower limb prosthesis alignment is crucial for optimal patient gait and function.
- Current alignment components are prohibitively expensive for low-income countries.
- There is a need for affordable prosthetic alignment solutions.
Purpose of the Study:
- To develop a lightweight and low-cost alignment coupler for lower limb prostheses.
- To create an accessible prosthetic fitting solution for resource-limited settings.
Main Methods:
- Designed and manufactured a reversible adjustable coupling.
- Recorded measurements of anterior/posterior, medial/lateral, and rotational adjustments.
- Performed mechanical testing, including repetitive loading, static, and strength tests.
- Assessed ease of use.
Main Results:
- The coupler allowed acceptable anterior/posterior and medial/lateral translation.
- It provided 30° of internal and external rotation.
- Mechanical tests, including 2000 cycles of repetitive loading, were successful.
- The component weighed 166g and cost $55.00 USD.
Conclusions:
- The reversible adjustable coupling offers adequate adjustment ranges for prosthesis alignment.
- It is suitable for both exoskeletal and endoskeletal prostheses.
- The low cost and weight make it an affordable option for low-income countries in clinical and educational settings.
More Related Videos
08:08Oscillation and Reaction Board Techniques for Estimating Inertial Properties of a Below-knee Prosthesis
Published on: May 8, 2014
08:42The DREAM Implant: A Lightweight, Modular, and Cost-Effective Implant System for Chronic Electrophysiology in Head-Fixed and Freely Behaving Mice
Published on: July 26, 2024
Related Concept Videos
The Anchoring-and-Adjustment Heuristic
Adjusting a Traverse
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and...
Components of Language
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Spin–Spin Coupling: Two-Bond Coupling (Geminal Coupling)
The central atom need not be NMR-active because its electrons are affected by the electron polarization of the spin-active atoms. However, spin information is transmitted less effectively than in one-bond coupling, and 2J values are usually weaker than 1J values. The energy of...