Development and reliability of a hand-held dynamometer device to measure the ankle dorsiflexor muscle strength
Masanobu Kimura1, Yuto Ogata1, Toru Akebi2
1University Hospital of Occupational and Environmental Health, Kitakyushu, Japan.
Background:
Ankle dorsiflexor muscle strength is a crucial component of gait.
Objective:
We describe the development of a simple, hand-held dynamometer to measure the ankle dorsiflexor muscle strength in the sitting position. In addition, we examine its intra- and inter-rater reliability.
Methods:
Measurements of the peak ankle dorsiflexor muscle strength were obtained by two examiners for 30 ankles of 15 healthy adults at two time points, with a one-day interval between measurements, to determine the inter- and intra-rater reliability. The intraclass correlation coefficients were calculated, and an intraclass correlation coefficient > 0.90 was considered as excellent reliability. A Bland-Altman analysis was used to assess systemic bias. The minimal detectable change in muscle strength was calculated with a confidence level of 95% (MDC95).
Results:
The reliability of the device was excellent for both intra- (intraclass correlation coefficients [1,3] = 0.94) and inter-rater (intraclass correlation coefficients [2,3] = 0.96) comparisons. No fixed or proportional bias was observed between the two examiners. The MDC95 was 0.77 N/kg.
Conclusions:
Our results indicate the excellent reliability and responsiveness of our device. By obtaining the measurements of dorsiflexor strength while sitting, compensatory motions are suppressed, yielding a more consistent measurement that can be reliably used to detect subtle changes in the ankle dorsiflexor muscle strength.
More Related Videos
12:59Improving Strength, Power, Muscle Aerobic Capacity, and Glucose Tolerance through Short-term Progressive Strength Training Among Elderly People
Published on: July 5, 2017
07:22Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
Published on: February 20, 2020
Related Concept Videos
Muscles of the Forearm that Move the Hand and Fingers
Anterior Compartment
The anterior compartment muscles originate from the humerus. They primarily function as flexors and are also known as flexor muscles. They typically insert on the carpals, metacarpals, and phalanges. The superficial layer includes the flexor carpi...
Reliability and Validity
Ankle Joint
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
The Equilibrium Binding Constant and Binding Strength
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...