Belt device for simplified CT-guided puncture and biopsy: a technical note
Cardiovascular and Interventional Radiology
|January 1, 1987
Summary
A novel radiolucent device enhances computed tomography (CT)-guided fine-needle punctures. This tool improves accuracy by allowing precise pre-setting of needle angle, depth, and position, simplifying procedures.
Area of Science:
- Medical Imaging
- Interventional Radiology
- Surgical Devices
Background:
- CT-guided fine-needle punctures are crucial for tissue sampling and interventions.
- Achieving optimal needle trajectory, depth, and position can be challenging, impacting accuracy and safety.
- Existing methods may require multiple adjustments or patient repositioning.
Purpose of the Study:
- To introduce and evaluate a new radiolucent device for CT-guided fine-needle punctures.
- To assess the device's capability in improving the precision of needle placement.
- To streamline the puncture procedure by enabling pre-setting and stationary patient positioning.
Main Methods:
- Development of a novel radiolucent device incorporating a needle holder sheath and goniometric scale.
- The device is designed as a belt for adjustable positioning around the patient.
- Pre-setting of needle parameters (angle, depth, position) based on CT monitor data.
Main Results:
- The device allows precise determination of the optimum angle, depth, and position for fine-needle punctures.
- Needle placement and control scans can be performed with the patient remaining stationary.
- The adjustable design facilitates versatile application during CT-guided procedures.
Conclusions:
- The new radiolucent device significantly enhances accuracy in CT-guided fine-needle punctures.
- It offers a streamlined approach, improving efficiency and potentially patient comfort.
- This innovation represents a valuable advancement in interventional radiology tools.