A fluorimetric nitrite biosensor with polythienothiophene-fullerene thin film detectors for on-site water monitoring
Nuno Miguel Matos Pires1, Tao Dong2, Zhaochu Yang3
1Chongqing Key Laboratory of Micro-Nano Systems and Smart Transduction, Collaborative Innovation Center on Micro-Nano Transduction and Intelligent Eco-Internet of Things, Chongqing Key Laboratory of Colleges and Universities on Micro-Nano Systems Technology and Smart Transducing, National Research Base of Intelligent Manufacturing Service, Chongqing Technology and Business University, Nan'an District, Chongqing 400067, China and Department of Microsystems - IMS, Faculty of Technology, Natural Sciences and Maritime Sciences, University of South-Eastern Norway-USN, P.O. Box 235, 3603 Kongsberg, Norway. tao.dong@usn.no and Centre for Environmental Radioactivity (CERAD CoE), Norwegian University of Life Sciences (NMBU), Faculty of Environmental Sciences and Natural Resource Management, P.O. Box 5003, NO-1432 Ås, Norway.
Abstract:
A novel fluorimetric sensor for highly sensitive nitrite detection on the site is presented in this study. The proposed on-chip approach comprises the use of integrated polymer photodetectors to detect light from fluorescence reactions with a diaminofluorescein probe. The detectors were prepared with a heterostructured nanofilm of polythieno[3,4-b]thiophene/benzodithiophene and (6,6)-phenyl-C71-butyric-acid methyl-ester as a photoactive layer. Prior to fluorimetric detection, the quality of the spin-coated photoactive layer was characterized via nano-morphology and current-density measurements. Nitrite assays were conducted on a poly(methyl methacrylate) microchannel chip, to which polythienothiophene-C71 based detectors were aligned. Results of signal-to-noise ratio determination have indicated a detection limit below 0.55 μM, lower than the 0.1 mg L-1 maximum limit of operation in recirculating aquaculture systems for farming Atlantic salmon Salmo salar. An increase of the nitrite concentration to toxic levels may therefore be possible to detect. The fluorimetric sensor exhibited good linearity over three orders of magnitude and acceptable detection reproducibility, which confirmed its analytical value. Further tests revealed great promise of the integrated biosensor device for detecting nitrite in aquaculture-relevant samples with high precision. The approach reported hereby may provide impetus to in situ analytical tools for monitoring water quality at aquaculture facilities, the food industries or water monitoring stations.
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors
States of Water
Water freezes when the intermolecular forces are greater than the kinetic energy. Unlike most other substances, water is less dense in its solid state than in its liquid state. This is because each water molecule can form...
Water and Mineral Acquisition