Related Experiment Video
Updated: Jan 23, 2026

Author Spotlight: Utilizing Venoplasty Balloon Model in Rodents to Simulate Surgical Interventions for Deep Veins
Published on: May 24, 2024
Randomized Controlled Study of an Absorbable Vena Cava Filter in a Porcine Model
Mitchell Eggers1, Serge Rousselle2, Mark Urtz3
1Adient Medical, 2315 Delta Bridge Dr, Pearland, TX 77854; University of Texas M.D. Anderson Cancer Center, Houston, Texas.
Purpose:
To compare the safety and efficacy of an absorbable inferior vena cava (IVC) filter and a benchmark IVC filter in a porcine model.
Materials And Methods:
A randomized controlled Good Laboratory Practice study was performed in Domestic Yorkshire cross swine. Sixteen swine were implanted with an absorbable IVC filter (test device; Adient Medical, Pearland, Texas); 8 were implanted with a benchmark metal IVC filter (control device; Cook Medical, Bloomington, Indiana). All animals underwent rotational digital subtraction pulmonary angiography and cavography (anteroposterior and lateral) before filter deployment and 5 and 32 weeks after deployment. Terminal procedures and necropsy were performed at 32 weeks. The IVC, heart, lungs, liver, and kidneys were harvested at necropsy. The reported randomized controlled GLP animal study was conducted at Synchrony Labs, Durham, North Carolina.
Results:
One animal died early in the test cohort of a recurring hemorrhage at the femoral access site resulting from a filter placement complication. All other animals remained clinically healthy throughout the study. No pulmonary embolism was detected at the 5- and 32-week follow-up visits. The absorbable filter subjects experienced less caval wall perforation (0% vs 100%) and thrombosis (0% vs 75%). The control device routinely perforated the IVC and occasionally produced collateral trauma to adjacent tissues (psoas muscle and aorta). The veins implanted with the absorbable filter were macroscopically indistinguishable from normal adjacent veins at 32 weeks except for the presence of radiopaque markers. Nontarget tissues showed no device-related changes.
Conclusions:
Implantation of the absorbable IVC filter in swine proved safe with no pulmonary emboli detected. There was complete to near-complete resorption of the filter polymer by 32 weeks with restoration of the normal appearance and structure of the IVC.
Related Concept Videos
Bioequivalence Experimental Study Designs: Completely Randomized and Randomized Block Designs
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Active Filters
Random Error
Random Variables
Uppercase letters such as X or Y denote a random variable. Lowercase letters like x or y denote the value of a random variable. If X is a random variable, then X is written in words, and x is given as a number.
For example, let X = the...