Estimated Average Glandular Dose for 1,828 Mammography Procedures in China: A Multicenter Study
Xiang DU1, Jin Wang1, Bao Li Zhu2
1Jiangsu Provincial Center for Disease Prevention and Control, Nanjing 210009, Jiangsu, China.
Objective:
To understand the distribution of the average glandular dose (AGD) in mammography by investigating 1,828 exposure parameters of 8 mammography machines in three cities, by using random sampling.
Methods:
A survey of 8 mammography machines in three different cities, sampled using stratified random sampling methods, was performed, and 1,828 mammography exposure parameters were recorded. Incident air kerma (k) was measured by Quality-Assurance (QA) dosimeters, and AGD was calculated by series conversion coefficients based on a 3D detailed Monte Carlo breast model, published by Wang et al. RESULTS: The distribution of compressed breast thickness (CBT) fitted a normal distribution, while that of AGD fitted a skewed distribution. The mean value of CBT in a medio-lateral oblique (MLO) view was about 5.6% higher than that in the craniocaudal (CC) view, with significant statistical difference; mean value of AGD and CBT in the sample was 1.3 mGy and 4.6 cm, respectively. The AGD trended upward with increasing CBT, similar to the results of other researches.
Conclusion:
The mean AGD and CBT levels in our study for mammography practice in China were 1.3 mGy and 4.6 cm, respectively. AGD is influenced by manufacturer-specific variation as machine response to CBT changes and target/filter combination. The present study can provide evidence for establishing a diagnostic reference level in China.
Related Concept Videos
Bioavailability Study Design: Single Versus Multiple Dose Studies
Average Acceleration
Average Power
Average Velocity
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Excretion: Pulmonary and Glandular Routes
Drugs can also be excreted in breast milk, which is crucial for breastfeeding infants. The...