Clinical validation of RIA-G, an automated optic nerve head analysis software
Digvijay Singh1, Srilathaa Gunasekaran2, Maya Hada3
1Noble Eye Care; Narayana Superspecialty Hospital, Gurugram, Haryana, India.
Purpose:
To clinically validate a new automated glaucoma diagnosis software RIA-G.
Methods:
A double-blinded study was conducted where 229 valid random fundus images were evaluated independently by RIA-G and three expert ophthalmologists. Optic nerve head parameters [vertical and horizontal cup-disc ratio (CDR) and neuroretinal rim (NRR) changes] were quantified. Disc damage likelihood scale (DDLS) staging and presence of glaucoma were noted. The software output was compared with consensus values of ophthalmologists.
Results:
Mean difference between the vertical CDR output by RIA-G and the ophthalmologists was - 0.004 ± 0.1. Good agreement and strong correlation existed between the two [interclass correlation coefficient (ICC) 0.79; r = 0.77, P < 0.005]. Mean difference for horizontal CDR was - 0.07 ± 0.13 with a moderate to strong agreement and correlation (ICC 0.48; r = 0.61, P < 0.05). Experts and RIA-G found a violation of the inferior-superior NRR in 47 and 54 images, respectively (Cohen's kappa = 0.56 ± 0.07). RIA-G accurately detected DDLS in 66.2% cases, while in 93.8% cases, output was within ± 1 stage (ICC 0.51). Sensitivity and specificity of RIA-G to diagnose glaucomatous neuropathy were 82.3% and 91.8%, respectively. Overall agreement between RIA-G and experts for glaucoma diagnosis was good (Cohen's kappa = 0.62 ± 0.07). Overall accuracy of RIA-G to detect glaucomatous neuropathy was 90.3%. A detection error rate of 5% was noted.
Conclusion:
RIA-G showed good agreement with the experts and proved to be a reliable software for detecting glaucomatous optic neuropathy. The ability to quantify optic nerve head parameters from simple fundus photographs will prove particularly useful in glaucoma screening, where no direct patient-doctor contact is established.
Related Concept Videos
Statistical Software for Data Analysis and Clinical Trials
Reliability and Validity
Muscles that Move the Head
The bilateral sternocleidomastoid, or SCM, and the suprahyoid and infrahyoid muscles are significant head flexors. The SCM muscles originate at the sternum and clavicle and attach to the mastoid process of the temporal bone. The SCM contracts bilaterally to bend the head forward, whereas...
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include:
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...