In-Vitro Pulsatile Flow Testing of Prosthetic Heart Valves: A Round-Robin Study by the ISO Cardiac Valves Working
Changfu Wu1, Neelakantan Saikrishnan2,3, Aaron J Chalekian4
1Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD, USA. Changfu.Wu@fda.hhs.gov.
A 2016 round-robin study found significant variability in prosthetic heart valve hydrodynamic testing among labs, despite improvements since 2005. This highlights the need for clearer ISO 5840 standards and better lab practices for accurate in vitro assessments.
Area of Science:
- Biomedical Engineering
- Medical Device Testing
- Cardiovascular Research
Background:
- Hydrodynamic performance testing is crucial for prosthetic heart valves per ISO 5840 standards.
- A 2005 round-robin study revealed significant inter-laboratory variability in these measurements.
- This variability necessitates re-evaluation of current testing protocols.
Purpose of the Study:
- To re-examine inter-laboratory variability in prosthetic heart valve hydrodynamic performance testing.
- To assess the state-of-the-art under the latest ISO 5840-1 and 5840-2:2015 standards.
- To identify areas for improvement in in vitro assessment procedures.
Main Methods:
- A 2016 round-robin study involving 13 international laboratories.
- Testing of St. Jude Medical Masters Series mechanical valves across various sizes.
- Standardized testing procedures with prespecified flow conditions.
Main Results:
- Improved, yet still significant, inter-laboratory variability was observed compared to the 2005 study.
- Coefficients of variation ranged from 7.7% to 21.6% for effective orifice area.
- Variability was also significant for regurgitant fraction (10.1-32.8%) and pressure gradient (14.7-45.5%).
Conclusions:
- The study identified ambiguities in current ISO 5840 standards and laboratory shortcomings.
- Clarifications are being incorporated into revised ISO 5840-1, -2, and -3 standards.
- Results can guide laboratories in benchmarking equipment and personnel training for in vitro assessments.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Anatomy of the Heart
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Physiology of the Heart: The Cardiac Cycle
Diastole: The Relaxation Phase
During diastole, all four heart chambers relax. The atrioventricular (AV) valves open, and the semilunar valves close. This phase sees the lowest chamber pressures, promoting ventricular filling. Venous blood enters the heart through the...