Determination of fecal alpha 1-antitrypsin concentration by radial immunodiffusion: two systems compared

C M Wilson1, K McGilligan, D W Thomas

  • 1Department of Pediatrics, Childrens Hospital of Los Angeles, CA 90027.

Clinical Chemistry
|February 1, 1988
PubMed

Insights

This study compared two radial immunodiffusion systems for measuring fecal alpha 1-antitrypsin (FA1AT). Both systems effectively measured FA1AT, though the Calbiochem system yielded higher values than the Helena system.

Area of Science:

  • Clinical Chemistry
  • Immunology
  • Biochemistry

Background:

  • Fecal alpha 1-antitrypsin (FA1AT) is a marker for protein-losing enteropathy.
  • Accurate measurement of FA1AT is crucial for diagnosing gastrointestinal disorders.

Purpose of the Study:

  • To compare the performance of Helena "QUIPlates" and Calbiochem "LC-Partigen" radial immunodiffusion systems for measuring fecal alpha 1-antitrypsin (FA1AT).
  • To establish reference ranges for FA1AT in different age groups and dietary conditions.

Main Methods:

  • Radial immunodiffusion assays were performed using both Helena "QUIPlates" and Calbiochem "LC-Partigen" systems.
  • Fecal samples from infants, children, and adults were analyzed.
  • Sample handling variations (storage, freeze-thaw, lyophilization) were investigated.

Main Results:

  • Calbiochem LC-Partigen system yielded FA1AT values approximately 30% higher than Helena QUIPlates.
  • Both systems successfully differentiated between normal and elevated FA1AT concentrations.
  • Sample storage conditions (room temperature, freeze-thaw, frozen storage) did not significantly impact FA1AT measurements.
  • Lyophilized and non-lyophilized samples showed poor correlation.

Conclusions:

  • Both Helena "QUIPlates" and Calbiochem "LC-Partigen" are suitable for measuring fecal alpha 1-antitrypsin (FA1AT).
  • Reference ranges should be established for each system due to observed differences in values.
  • Standard sample handling procedures are reliable, but lyophilization may affect results.