Cochlear Implant Surgery and Electrically-evoked Auditory Brainstem Response Recordings in C57BL/6 Mice
Published on: January 9, 2019
Examination and Comparison of Electrically Evoked Compound Action Potentials and Electrically Evoked Auditory
Seda Bayrak1, Başak Mutlu2, Günay Kırkım1
1Department of Audiology, Dokuz Eylül University School of Health Sciences, İzmir, Turkey.
Objective:
To investigate the relationship between electrically evoked compound action potentials (ECAP) and electrically evoked auditory brainstem responses (EABR) in children with cochlear implants (CI) without inner ear anomalies.
Methods:
Sixteen children between the ages of two and six years who were CI users participated in the study. ECAP thresholds were recorded from one electrode in the basal, medial, and apical regions of the cochlear implant. EABRs were recorded from electrodes whose ECAP thresholds were determined. The latency-intensity functions, amplitude and morphological analyzes of the eIII and eV waves at 200 and 180 current unit (CU) excitation levels were performed. The data obtained were analyzed statistically.
Results:
ECAP thresholds were found to be 171.5±11.38, 169.69±20.32 and 160.81±20.03 CU at the basal, medial and apical electrodes, respectively. EABR thresholds were also found to be 169.69±12.17, 165.62±16.41 and 160±15.49 CU in basal, medial and apical electrodes, respectively. There was a strong positive correlation between ECAP and EABR thresholds in apical, medial and basal electrodes (p<0.05). EABR threshold levels were not significantly different between basal, medial and apical region electrodes (p>0.05), and ECAP threshold values were significantly different between apical and basal region electrodes (p=0.002). When the significance values of EABR eV wave latencies were analyzed in terms of electrode region, the difference between basal and apical regions was found to be significant (p=0.03).
Conclusion:
Consistency was found between ECAP and EABR recordings. However, it was concluded that one could not be preferred over the other because the data quality of the two tests was different. In future studies, ECAP and EABR recordings may be recommended by selecting more electrodes for stimulation.
Related Concept Videos
Finding Electric Potential From Electric Field
Determining Electric Field From Electric Potential
In general, regardless of whether the electric field is uniform, it points in the direction of decreasing potential because the force on a positive...
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Electric Potential and Potential Difference
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...
Electric Potential Energy in a Uniform Electric Field
Calculations of Electric Potential I
The ring is divided into infinitesimal small arcs such that point M is equidistant from all the arcs. Here, the cylindrical coordinate system is used to calculate the electric potential at point M. A general element of the arc between angles θ and θ + dθ is of the length Rdθ and has a charge of...