Collodion Remover Can Degrade Plastic-Containing Medical Devices Commonly Used in the Intensive Care Unit
Mark P Fitzgerald1, Maureen Donnelly1, Lisa Vala1
1Division of Neurology, Departments of Neurology and Pediatrics, Children's Hospital of Philadelphia and the Perelman School of Medicine at the University of Pennsylvania , Philadelphia , Pennsylvania.
Abstract:
Collodion remover, a solvent blend used to remove collodion glue after long-term video EEG monitoring, was implicated as a potential causative factor in patient safety events at our institution during which damage to plastic components of medical devices was noted in the intensive care unit. We sought to determine experimentally whether collodion remover could lead to degradation of multiple plastic-containing medical devices commonly used in the intensive care unit to determine whether workflow changes were needed during electrode removal. We exposed devices to collodion remover for brief, intermediate, and prolonged durations. We report that collodion remover is capable of degrading the hard plastic components of multiple medical devices after prolonged exposure; however, intermediate duration exposure was also capable of producing damage to clave connectors used with intravenous and central lines, which could plausibly lead to adverse events given the widespread use of these devices. These data suggest a pathway-based approach to collodion remover use might be beneficial in minimizing the potential impact of this solvent on plastic-containing medical devices.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
Related Concept Videos
Measurement: Derived Units
Measurement: Standard Units
Regulated Protein Degradation
Protein degradation plays two important roles in the cells. It helps to protect cells from misfolded or damaged proteins before they lead to a...
Proteins: From Genes to Degradation
Transcription is the synthesis of RNA...
Common Ion Effect
Plasticizers
Plasticizers function by using surface-active agents to create repulsive electrostatic forces between cement particles. This dispersion enhances the concrete's...