[In vitro experimental study on the mechanical properties of biodegradable polymer stents]
Yunbo Wei1, Minjie Wang2, Danyang Zhao1
1School of Mechanical Engineering, Dalian University of Technology, Dalian, Liaoning 116024, P.R.China.
Abstract:
In vitro experimental test for mechanical properties of a vascular stent is a main method to evaluate its effectiveness and safety, which is of great significance to the clinical applications. In this study, a comparative study of planar, V-groove and radial compression methods for the radial support property test were performed, and the effects of compression rate and circumferential position on the test results were conducted. Based on the three-point bending method, the influences of compression rate and circumferential position on flexibility were also explored. And then a best test proposal was selected to evaluate the radial support property and flexibility of the three self-designed stents and the comparative biodegradable vascular stent (BVS) (BVS1.1, Abbott Vascular, USA) with different outside diameters of 1.4 mm, 1.7 mm and 2.4 mm. The results show that the developing trends of the compression load with the compression displacement measured by the three radial support property test methods are the same, but normalized radial force values are quite different. The planar compression method is more suitable for comparing the radial support properties of stents with different diameters and structures. Compression rate has no obvious effect on the testing results of both the radial support property and flexibility. Compression circumferential position has a great impact on testing radial support property with the planar or V-groove compression methods and testing flexibility with three-point bending method. The radial support properties of all the three self-designed stents are improved at a certain degree compared to that of the BVS stent. The study has better guide significance and reference value for testing mechanical properties of vascular stents.
More Related Videos
11:07Biodegradable Magnesium Stent Treatment of Saccular Aneurysms in a Rat Model - Introduction of the Surgical Technique
Published on: October 1, 2017
07:15A Testing Platform for Durability Studies of Polymers and Fiber-reinforced Polymer Composites under Concurrent Hygrothermo-mechanical Stimuli
Published on: December 11, 2014
Related Concept Videos
Polymers
Polymers
Reaction Mechanisms
For instance, the decomposition of ozone appears to follow a mechanism with two steps:
Physical and Chemical Properties of Matter
Properties of Transition Metals
Methods for Studying Drug Absorption: In vitro
The diffusion cell method uses a two-compartment cell, including a donor compartment with the drug solution, which simulates the environment where the drug is applied, and a receptor compartment with a buffer solution, which simulates the environment...