Quantitative enhancement thresholds and machine learning algorithms for the evaluation of renal lesions using
Markus M Obmann1,2, Aurelio Cosentino3, Joshy Cyriac4
1Clinic of Radiology and Nuclear Medicine, University Hospital Basel, University of Basel, Petersgraben 4, 4031, Basel, Switzerland. Markus.Obmann@usb.ch.
Purpose:
To establish thresholds for contrast enhancement-based attenuation (CM) and iodine concentration (IOD) for the quantitative evaluation of enhancement in renal lesions on single-phase split-filter dual-energy CT (tbDECT) and combine measurements in a machine learning algorithm to potentially improve performance.
Material:
126 patients with incidental renal cysts (both hypo- and hyperdense cysts) or high suspicion for renal cell carcinoma (312 total lesions) undergoing abdominal, portal venous phase tbDECT were initially included in this retrospective study. Gold standard was pathological confirmation or follow-up imaging (MRI or multiphasic CT). CM, IOD, and ROI size were recorded. Thresholds for CM and IOD were identified using Youden-Index of the empirical ROC curves. Decision tree (DTC) and random forest classifier (RFC) were trained. Sensitivities, specificities, and AUCs were compared using McNemar and DeLong test.
Results:
The final study cohort comprised 40 enhancing and 113 non-enhancing renal lesions. Optimal thresholds for quantitative iodine measurements and contrast enhancement-based attenuation were 1.0 ± 0.0 mg/ml and 23.6 ± 0.3 HU, respectively. Single DECT parameters (IOD, CM) showed similar overall performance with an AUC of 0.894 and 0.858 (p = 0.541) (sensitivity 90 and 80%, specificity 88 and 92%, respectively). While overall performance for the DTC (AUC 0.944) was higher than RFC (AUC 0.886), this difference (p = 0.409) and comparison to CM (p = 0.243) and IOD (p = 0.353) was not statistically significant.
Conclusions:
Enhancement in incidental renal lesions on single-phase tbDECT can be classified with up to 87.5% sensitivity and 94.6% specificity. Algorithms combining DECT parameters did not increase overall performance.
More Related Videos
07:15Machine Learning Algorithms for Early Detection of Bone Metastases in an Experimental Rat Model
Published on: August 16, 2020
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Inductance: Single-Phase And Three-Phase Line
Single-Phase Two-Wire Line:
A single-phase line consists of two solid cylindrical conductors, denoted as x and y. Each conductor carries phasor currents ix and iy, respectively. Given that the sum of these currents is...
Capacitance: Single-Phase And Three-Phase Line
Single-Phase Lines
Consider a single-phase, two-wire transmission line with equal phase spacing energized by a voltage source. One conductor carries a uniform positive charge, while the other carries an equal negative charge. The capacitance C of the line can be derived from the voltage V between the conductors. For a one-meter section of the line, the capacitance is given...
Three-Phase Short Circuit—Unloaded Synchronous Machine
This behavior occurs due to the magnetic flux produced by the short-circuit armature currents. Initially, these currents follow high-reluctance paths but eventually shift to...
Power Distribution in Three-phase and Single Phase Circuits
Single-Phase Power Distribution:
Single-phase circuits are typical in household settings;...
Energy-requiring Steps of Glycolysis