Objective:
The objective of this study was to evaluate a prototype, ultrahigh-resolution computed tomography offering higher reconstruction matrix (1024 × 1024) and spatial resolution (0.15 mm) for chest imaging.
Methods:
Higher (1024) matrix reconstruction enabled by ultrahigh-resolution computed tomography scanner (128-detector rows; detector width, 0.25 mm; spatial resolution, 0.15 mm) was compared with conventional (512) reconstruction with image quality grading on a Likert scale (1, excellent; 5, nondiagnostic) for image noise, artifacts, contrast, small detail, lesion conspicuity, image sharpness, and diagnostic confidence. Image noise and signal-to-noise ratio were quantified.
Results:
Diagnostic image quality was achieved for all scans on 101 patients. The 1024 reconstruction demonstrated increased image noise (20.2 ± 4.0 vs 17.2 ± 3.8, P < 0.001) and a worse noise rating (1.98 ± 0.63 vs 1.75 ± 0.61, P < 0.001) but performed significantly better than conventional 512 matrix with fewer artifacts (1.37 ± 0.43 vs 1.50 ± 0.48, P < 0.001), better contrast (1.50 ± 0.56 vs 1.62 ± 0.57, P < 0.001), small detail detection (1.06 ± 0.19 vs 2.02 ± 0.22, P < 0.001), lesion conspicuity (1.08 ± 0.23 vs 2.02 ± 0.24, P < 0.001), sharpness (1.09 ± 0.24 vs 2.02 ± 0.28, P < 0.001), and overall diagnostic confidence (1.09 ± 0.25 vs 1.18 ± 0.34, P < 0.001).
Conclusions:
Ultrahigh-resolution computed tomography enabled a higher reconstruction matrix and improved image quality compared with conventional matrix reconstruction, with a minor increase in noise.
More Related Videos
Related Concept Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
06:31Erosion Identification in Metacarpophalangeal Joints in Rheumatoid Arthritis using High-Resolution Peripheral Quantitative Computed Tomography
04:24Proper Positioning and Restraint of a Rat Hind Limb for Focused High Resolution Imaging of Bone Micro-architecture Using In Vivo Micro-computed Tomography
Visualization of a Cerebral Thrombus in a Mouse Using Micro-Computed Tomography Imaging
Imaging Studies III: Computed Tomography