Objective:
Lead orientation is a new degree of freedom with directional deep brain stimulation (DBS) leads. We investigated how prevalent deviations from the intended implantation direction are in a large patient cohort.
Methods:
The Directional Orientation Detection (DiODe) algorithm to determine lead orientation from postoperative CT scans was implemented into the open-source Lead-DBS toolbox. Lead orientation was analyzed in 100 consecutive patients (198 leads). Different anatomical targets and intraoperative setups were compared.
Results:
Deviations of up to 90° from the intended implantation direction were observed. Deviations of more than 30° were seen in 42% of the leads and deviations of more than 60° in about 11% of the leads. Deviations were independent from the neuroanatomical target and the stereotactic frame but increased depending on which microdrive was used.
Discussion:
Our results indicate that large deviations from the intended implantation direction are a common phenomenon in directional leads. Postoperative determination of lead orientation is thus mandatory for investigating directional DBS.
Related Concept Videos
08:06Robotic Cochlear Implantation for Direct Cochlear Access
14:49Direct and Indirect Culture Methods for Studying Biodegradable Implant Materials In Vitro
Standard Deviation
Mean Absolute Deviation
Let us consider a dataset containing the number of unsold cupcakes in five shops: 10, 15, 8, 7, and 10. Initially, calculate the sample mean. Then calculate the deviation, or the difference, between each data value and the mean. Next, the absolute values of these deviations are added and divided by the sample size to...
Variation: Normal Distribution, Range, and Standard Deviation
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function