Experimental and Numerical Simulation of Biodegradable Stents with Different Strut Geometries
Chong Chen1, Yan Xiong2, Wentao Jiang3
1School of Mechanical Engineering, Sichuan University, Chengdu, 610065, China.
Objective:
Biodegradable stents represent a new technological development in the field of cardiovascular angioplasty. The stent geometry plays a crucial role in determining stent performance.
Methods:
This study describes four stent models with various strut geometries (circular, triangular, hexagonal, and spline) and identical links. The performance of the various stents is assessed by modeling the processes of compression and balloon expansion using experimental and numerical techniques.
Results:
Four adjustable variables related to radial strength, foreshortening rate, maximum expanded diameter and coverage are considered in this studies. The maximum stress distributions from our numerical analysis are in agreement with the experimental fracture measurements.
Conclusion:
Analysis of the stent parameters indicates that the hexagonal geometry is a relatively good stent strut design. The results described in this paper will be useful in further optimization studies and for the development of stents with a higher radial strength and increased fracture resistance.