Oxygenator Impact on Ceftolozane and Tazobactam in Extracorporeal Membrane Oxygenation Circuits
Jeffrey J Cies, Wayne S Moore1, Nadji Giliam2
1The Center for Pediatric Pharmacotherapy LLC, Pottstown, PA.
Objectives:
To determine the oxygenator impact on alterations of ceftolozane/tazobactam in a contemporary neonatal/pediatric (1/4-inch) and adolescent/adult (3/8-inch) extracorporeal membrane oxygenation circuit including the Quadrox-i oxygenator (Maquet, Wayne, NJ).
Design:
A 1/4-inch and 3/8-inch, simulated closed-loop extracorporeal membrane oxygenation circuits were prepared with a Quadrox-i pediatric and Quadrox-i adult oxygenator and blood primed. Additionally, 1/4-inch and 3/8-inch circuits were also prepared without an oxygenator in series. A one-time dose of ceftolozane/tazobactam was administered into the circuits and serial preoxygenator and postoxygenator concentrations were obtained at 5 minutes, 1, 2, 3, 4, 5, 6, and 24-hour time points. Ceftolozane/tazobactam was also maintained in a glass vial and samples were taken from the vial at the same time periods for control purposes to assess for spontaneous drug degradation SETTING:: A free-standing extracorporeal membrane oxygenation circuit.
Patients:
None.
Interventions:
Single-dose administration of ceftolozane/tazobactam into closed-loop extracorporeal membrane oxygenation circuits prepared with and without an oxygenator in series with serial preoxygenator, postoxygenator, and reference samples obtained for concentration determination over a 24-hour study period.
Measurements And Main Results:
For the 1/4-inch circuit, there was approximately 92% ceftolozane and 22-25% tazobactam loss with the oxygenator in series and 19-30% ceftolozane and 31-34% tazobactam loss without an oxygenator in series at 24 hours. For the 3/8-inch circuit, there was approximately 85% ceftolozane and 29% tazobactam loss with the oxygenator in series and 25-27% ceftolozane and 23-26% tazobactam loss without an oxygenator in series at 24 hours. The reference ceftolozane and tazobactam concentrations remained relatively constant during the entire study period demonstrating the drug loss in each size of the extracorporeal membrane oxygenation circuit with or without an oxygenator was not a result of spontaneous drug degradation.
Conclusions:
This ex vivo investigation demonstrated substantial ceftolozane loss within an extracorporeal membrane oxygenation circuit with an oxygenator in series with both sizes of the Quadrox-i oxygenator at 24 hours and significant ceftolozane loss in the absence of an oxygenator. Tazobactam loss was similar regardless of the presence of an oxygenator. Further evaluations with multiple dose in vitro and in vivo investigations are needed before specific drug dosing recommendations can be made for clinical application with extracorporeal membrane oxygenation.
More Related Videos
07:38Design and Development of a Model to Study the Effect of Supplemental Oxygen on the Cystic Fibrosis Airway Microbiome
Published on: August 3, 2021
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Transport in the Blood
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: