Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
An experimental model of veno-venous arterial extracorporeal membrane oxygenation
Mirko Belliato1, Luca Caneva1, Alessandro Aina2
1UOC Anestesia e Rianimazione, IRCCS Policlinico San Matteo, Pavia, Italy.
Introduction:
Veno-venous arterial extracorporeal membrane oxygenation is a hybrid-modality of extracorporeal membrane oxygenation combining veno-venous and veno-arterial extracorporeal membrane oxygenation. It may be applied to patients with both respiratory and cardio-circulatory failure.
Aim:
To describe a computational spreadsheet regarding an ex vivo experimental model of veno-venous arterial extracorporeal membrane oxygenation to determine the return of cannula pairs in a single pump-driven circuit.
Methods:
We developed an ex vivo model of veno-venous arterial extracorporeal membrane oxygenation with a single pump and two outflow cannulas, and a glucose solution was used to mimic the features of blood. We maintained a fixed aortic impedance and physiological pulmonary resistance. Both flow and pressure data were collected while testing different pairs of outflow cannulas. Six simulations of different cannula pairs were performed, and data were analysed by a custom-made spreadsheet, which was able to predict the flow partition at different flow levels.
Results:
In all simulations, the flow in the arterial cannula gradually increased differently depending on the cannula pair. The best cannula pair was a 19-Fr/18-cm arterial with a 17-Fr/50-cm venous cannula, where we observed an equal flow split and acceptable flow into the arterial cannula at a lower flow rate of 4 L/min.
Conclusion:
Our computational spreadsheet identifies the suitable cannula pairing set for correctly splitting the outlet blood flow into the arterial and venous return cannulas in a veno-venous arterial extracorporeal membrane oxygenation configuration without the use of external throttles. Several limitations were reported regarding fixed aortic impedance, central venous pressure and the types of cannulas tested; therefore, further studies are mandatory to confirm our findings.
Related Concept Videos
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Two-Compartment Open Model: Extravascular Administration
The absorption exponent (ka) indicates the speed at which the drug...
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...