Validation of Highly Accelerated Wave-CAIPI SWI Compared with Conventional SWI and T2*-Weighted Gradient
J Conklin1, M G F Longo2, S F Cauley2,3
1From the Department of Radiology (J.C., M.G.F.L., S.F.C., K.S., R.G.G., P.W.S., J.E.K., O.R., S.Y.H.), Massachusetts General Hospital, Boston, Massachusetts jconklin1@mgh.harvard.edu.
AJNR. American Journal of Neuroradiology
|November 16, 2019
Summary
Accelerated wave-controlled aliasing in parallel imaging (CAIPI) SWI offers faster brain imaging. This advanced SWI technique provides superior visualization of pathology and reduced motion artifacts compared to conventional methods.
Area of Science:
- Radiology
- Medical Imaging
- Neuroimaging
Background:
- Susceptibility-weighted imaging (SWI) is crucial for detecting intracranial hemorrhage and mineralization.
- Standard SWI sequences often involve long acquisition times, limiting their clinical utility.
- T2*-weighted gradient recalled-echo (T2*W GRE) is a common alternative for brain imaging.
Purpose of the Study:
- To compare a highly accelerated wave-controlled aliasing in parallel imaging (CAIPI) SWI sequence with standard SWI and T2*W GRE for routine 3T brain imaging.
- To evaluate the diagnostic performance, speed, and artifact levels of wave-CAIPI SWI.
- To assess the superiority of wave-CAIPI SWI over T2*W GRE and its noninferiority to standard SWI.
Main Methods:
- Prospective evaluation of 246 adult patients using conventional SWI/T2*W GRE and wave-CAIPI SWI.
- Wave-CAIPI SWI acquisition was 3-5 times faster than standard SWI.
- Two blinded radiologists assessed hemorrhage presence, microhemorrhage count, and motion artifact severity using forced-choice comparisons.
Main Results:
- Wave-CAIPI SWI detected significantly more hemorrhages and microhemorrhages than T2*W GRE (P < .001).
- Wave-CAIPI SWI demonstrated superior visualization of pathology, artifacts, and overall diagnostic quality compared to T2*W GRE (all P < .001).
- Wave-CAIPI SWI was noninferior to standard SWI for pathology visualization, artifact assessment, and diagnostic quality, with significantly reduced motion artifacts (P < .01).
Conclusions:
- Accelerated wave-CAIPI SWI provides superior diagnostic quality and pathology visualization compared to T2*W GRE.
- Wave-CAIPI SWI is noninferior to standard SWI, offering reduced scan times and motion artifacts.
- This advanced SWI technique enhances routine clinical brain imaging at 3T.
Related Concept Videos
Magnetic Resonance Imaging
8.9K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
8.9K
Imaging Studies for Cardiovascular System IV: CMRI
284
Cardiovascular magnetic resonance imaging, or CMRI, is a non-invasive diagnostic test that employs a magnetic field and radiofrequency waves to create precise images of the heart and arteries. It provides comprehensive information about cardiac anatomy, function, perfusion, and tissue characterization without ionizing radiation.IndicationsCMRI diagnoses various heart conditions, including tissue damage from heart attacks, ischemic heart disease, myocarditis, aortic issues (tears, aneurysms,...