Method to calculate frequency characteristics of reconstruction filter kernel in X-ray computed tomography
Kazuhiro Sato1, Yu Tomita2, Ryota Kageyama3
1Tohoku University Graduate School of Medicine, 2-1 Seiryo-machi, Aoba-ku, Sendai, Miyagi, 980-8575, Japan. satok@med.tohoku.ac.jp.
Abstract:
A computed tomography (CT) image is generally reconstructed by a filtered back projection (FBP) algorithm. In an FBP algorithm, the image quality primarily depends on a reconstruction filter kernel. Although the details of the filter kernel are not disclosed to users, the frequency response of the filter kernel can theoretically be calculated using the relational formula of the filter kernel and the modulation transfer function (MTF) of the reconstruction algorithm (MTFA). In this study, we proposed a method to determine the frequency response of a filter kernel and verify its validity. Two clinical CT scanners were used to derive the filter kernel. The MTF was obtained and subsequently separated to the MTF of the scanner system and MTFA. Using the relational formula of the filter kernel and MTFA, we calculated the frequency response of the filter kernel. To verify the calculated result, we measured the noise power spectrum (NPS). Additionally, the filter kernel was calculated using the relational formula of the filter kernel and NPS. In both CT scanners, the filter kernels calculated by the two methods showed good agreement, and we confirmed the validity of the results and the effectiveness of the proposed method. Furthermore, the inherent image quality performance of the CT scanner could be clarified by the reconstruction filter kernel.
More Related Videos
09:00Visualization of Failure and the Associated Grain-Scale Mechanical Behavior of Granular Soils under Shear using Synchrotron X-Ray Micro-Tomography
Published on: September 29, 2019
10:16X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
Published on: August 20, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Aliasing
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original...
Reconstruction of Signal using Interpolation
Imaging Studies III: Computed Tomography
Fast Fourier Transform
The computational efficiency of the FFT becomes...
Sampling Theorem