Pilot study: Internally cooled orthopedic drills - standard sterilization is not enough?
Tomislav Bruketa1, Goran Augustin1, Selma Pintarić1
11Department of Surgery, Zagreb University Hospital Centre and School of Medicine, University of Zagreb, Zagreb, Croatia; 2Department of Microbiology and Infectious Diseases with Clinic, Veterinary Faculty, University of Zagreb, Zagreb, Croatia; 3Trauma and Orthopaedics Department, Brighton and Sussex University Hospitals, NHS Trust, Brighton, United Kingdom.
Standard sterilization of internally cooled bone drills may be insufficient. Extended autoclave times of 20 minutes or more are recommended to ensure sterility and prevent thermal osteonecrosis risk.
Area of Science:
- Medical Devices
- Sterilization Science
- Orthopedic Surgery
Background:
- Bone drilling generates heat, potentially causing thermal osteonecrosis.
- Internally cooled bone drills reduce heat but may retain bacteria in lumens.
- Standard sterilization protocols might be inadequate for these novel drills.
Purpose of the Study:
- To evaluate the effectiveness of autoclaving for internally cooled, open-type bone drills.
- To determine optimal sterilization parameters for these medical devices.
- To provide evidence-based recommendations for sterilizing lumened bone drills.
Main Methods:
- Internally cooled bone drills were artificially contaminated with various bacteria.
- Drills were subjected to autoclaving at 132°C and 2.6 bar for 15, 20, and 30 minutes.
- Post-sterilization bacterial presence was assessed to determine effectiveness.
Main Results:
- A 15-minute sterilization cycle showed a 93.75% sterility rate, with one drill remaining contaminated.
- Extended autoclave cycles of 20 and 30 minutes achieved 100% sterility for all tested drills.
- No bacterial contamination was detected after 20 or 30 minutes of autoclaving.
Conclusions:
- Standard autoclave times may not guarantee sterility for internally cooled bone drills.
- An extended sterilization time of at least 20 minutes is minimally recommended.
- Ensuring complete sterility is crucial for preventing surgical site infections and complications.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Key Techniques in Microbiology