Related Experiment Video
Updated: Jan 2, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Enantiomeric Separation of Meclizine Hydrochloride in Pharmaceutical Dosage Form by HPLC Method
Byran Gowramma1, Ramachandran Senthil Kumar2, Rajagopal Kalirajan1
1Department of Pharmaceutical Chemistry, JSS College of Pharmacy, (JSS Academy of Higher Education and Research), Udhagamandalam, Nilgiris-643001, Tamilnadu, India.
Objective:
A basic, powerful and isocratic chiral fluid chromatographic technique was created and approved for the enantiomeric partition of meclizine hydrochloride in pharmaceutical dose structure.
Methods:
The chromatographic partition was accomplished on Phenomenex® Lux Cellulose 1 (250 mm x 4.6 mm i.d, 5 μm molecule size) section utilizing portable stage framework containing acetonitrile: 25mM ammonium bicarbonate (75:25%v /v). The versatile stage was siphoned on the segment at the stream pace of 1.0 mL/min, and UV recognition was done at 230 nm.
Result:
The breaking points of recognition and measurement were observed to be 0.25 μg/mL and 1.00 μg/mL individually, for 20μL infusion volume. The alignment bend demonstrated phenomenal linearity over the focus scope of 1-5 μg/mL for (±) meclizine enantiomers with a relationship coefficient (r2 = 0.999). The recuperation investigation of meclizine from tablet plan was observed to be 97.33% and 98.81% separately. Meclizine standard arrangement and versatile stage were observed to be steady for in any event 32h. The meclizine enantiomers were very much settled with mean maintenance times of about (+) Meclizine at 13.14 min and (-) Meclizine at 14.33 min individually.
Conclusion:
The created technique was broadly approved and demonstrated to be hearty, exact, exact and appropriate for the examination of meclizine enantiomers in tablet measurement structure and security investigations of meclizine.
More Related Videos
11:53An Alternative Culture Method to Maintain Genomic Hypomethylation of Mouse Embryonic Stem Cells Using MEK Inhibitor PD0325901 and Vitamin C
Published on: June 1, 2018
08:57Sample Extraction and Simultaneous Chromatographic Quantitation of Doxorubicin and Mitomycin C Following Drug Combination Delivery in Nanoparticles to Tumor-bearing Mice
Published on: October 5, 2017
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
Racemic Mixtures and the Resolution of Enantiomers
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...