Uncertainty contribution of derivatization in gas chromatography/mass spectrometric analysis
Martin Vilbaste1, Eliise Tammekivi1, Ivo Leito1
1University of Tartu, Institute of Chemistry, Ravila 14a, 50411, Tartu, Estonia.
Rationale:
The purpose of the current work is to realistically assess the uncertainty contribution in gas chromatography/mass spectrometry (GC/MS) analysis originating from less-than-ideal derivatization efficiency.
Methods:
As the exemplary analytical method a two-step derivatization method with KOH and BSTFA (N,O-bis(trimethylsilyl)trifluoroacetamide), applied for the analysis of fatty acid triglycerides (using real measurement data), was selected. The derivatization efficiencies were in the range 0.89-1.04. In this study, two approaches for bottom-up uncertainty evaluation were compared: the traditional GUM approach and the Monte Carlo method (MCM). Both were used with and without taking correlation between input quantities into account.
Results:
The most reliable uncertainty estimates were in the range 0.07-0.08 (expanded uncertainties at 95% coverage probability). A strong negative correlation was found between the slope and intercept of the calibration graph (r = -0.71) and it markedly influenced the uncertainty estimate of derivatization efficiency. The MCM was found to give somewhat higher uncertainty estimates, which are considered more realistic.
Conclusions:
Derivatization directly affects the analysis result. Thus, in the case of this exemplary analysis, just derivatization alone (i.e. if all other uncertainty sources are neglected) causes relative expanded uncertainty around 8%, being thus an important and in some cases the dominant uncertainty contributor.
More Related Videos
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Propagation of Uncertainty from Systematic Error
Uncertainty: Overview
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Mass Spectrum: Interpretation